Related Experiment Video
Updated: May 5, 2026

Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
Au Nanoshell-Based Lateral Flow Immunoassay for Colorimetric and Photothermal Dual-Mode Detection of Interleukin-6
Congying Wen1, Yue Dou1, Yao Liu1
1College of Chemistry and Chemical Engineering, China University of Petroleum (East China), Qingdao 266580, China.
Insights
This study introduces Aushell nanoparticles for enhanced Interleukin-6 (IL-6) detection using lateral flow immunoassay (LFIA). The novel dual-mode detection significantly improves sensitivity and quantitative accuracy for disease monitoring.
Area of Science:
- Nanotechnology
- Biomedical Engineering
- Analytical Chemistry
Background:
- Interleukin-6 (IL-6) is a critical biomarker for disease progression and therapeutic efficacy.
- Current lateral flow immunoassay (LFIA) methods for IL-6 detection lack sensitivity and quantitative precision.
- Point-of-care testing (POCT) requires sensitive and reliable detection methods for IL-6.
Purpose of the Study:
- To develop a novel LFIA reporter using Aushell nanoparticles for dual-mode IL-6 detection.
- To enhance the sensitivity and quantitative ability of IL-6 detection in LFIA.
- To assess the practical application potential of Aushell-based LFIA for IL-6 POCT.
Main Methods:
- Synthesis of Aushell nanoparticles via galvanic exchange, tuning localized surface plasmon resonance (LSPR) to the near-infrared (NIR) range.
- Modification of Aushell nanoparticles with IL-6 detection antibodies to create Aushell probes.
- Development of a dual-mode LFIA combining colorimetric (naked-eye) and photothermal (infrared thermal camera) detection.
Main Results:
- Colorimetric detection achieved a limit of 5 ng/mL for IL-6.
- Photothermal detection enabled quantitative measurement of IL-6 from 1-1000 ng/mL with a detection limit of 0.3 ng/mL.
- The Aushell-based LFIA demonstrated improved sensitivity (nearly 20-fold reduction in detection limit) and quantitative ability compared to traditional methods.
Conclusions:
- Aushell nanoparticles serve as effective reporters for dual-mode LFIA, significantly enhancing IL-6 detection.
- The developed method offers high sensitivity, quantitative accuracy, specificity, and retains the advantages of LFIA (convenience, speed, cost-effectiveness).
- This Aushell-based LFIA shows strong potential for practical IL-6 POCT applications.
Abstract:
Interleukin-6 (IL-6) detection and monitoring are of great significance for evaluating the progression of many diseases and their therapeutic efficacy. Lateral flow immunoassay (LFIA) is one of the most promising point-of-care testing (POCT) methods, yet suffers from low sensitivity and poor quantitative ability, which greatly limits its application in IL-6 detection. Hence, in this work, we integrated Aushell nanoparticles (NPs) as new LFIA reporters and achieved the colorimetric and photothermal dual-mode detection of IL-6. Aushell NPs were conveniently prepared using a galvanic exchange process. By controlling the shell thickness, their localized surface plasmon resonance (LSPR) peak was easily tuned to near-infrared (NIR) range, which matched well with the NIR irradiation light. Thus, the Aushell NPs were endowed with good photothermal effect. Aushell NPs were then modified with IL-6 detection antibody to construct Aushell probes. In the LFIA detection, the Aushell probes were combined with IL-6, which were further captured by the capture IL-6 antibody on the test line of the strip, forming a colored band. By observation with naked eyes, the colorimetric qualitative detection of IL-6 was achieved with limit of 5 ng/mL. By measuring the temperature rise of the test line with a portable infrared thermal camera, the photothermal quantitative detection of IL-6 was performed from 1~1000 ng/mL. The photothermal detection limit reached 0.3 ng/mL, which was reduced by nearly 20 times compared with naked-eye detection. Therefore, this Aushell-based LFIA efficiently improved the sensitivity and quantitative ability of commercial colloidal gold LFIA. Furthermore, this method showed good specificity, and kept the advantages of convenience, speed, cost-effectiveness, and portability. Therefore, this Aushell-based LFIA exhibits practical application potential in IL-6 POCT detection.