Performance Evaluation of Open Channel Buhlmann Fecal Calprotectin Turbo Assay on Abbott Alinity C Analyzer

Kavithalakshmi Sataranatarajan1, Shishir Adhikari2, Ngoc Nguyen2

  • 1Department of Pathology, University of Texas Southwestern Medical Center, Dallas, TX 75235, USA.

PubMed

Insights

A new automated fecal calprotectin (fCAL) assay (AFCAL) shows excellent performance for diagnosing inflammatory bowel disease (IBD). This validated method improves efficiency and reduces costs in clinical laboratories.

Area of Science:

  • Gastroenterology
  • Clinical Chemistry
  • Laboratory Medicine

Background:

  • Inflammatory bowel disease (IBD) involves chronic gastrointestinal inflammation.
  • Fecal calprotectin (fCAL) is a key noninvasive biomarker for IBD diagnosis and monitoring.
  • Current fCAL testing relies on ELISA methods, prompting the need for automated solutions.

Purpose of the Study:

  • To validate the performance of the Buhlmann fCAL turbo assay on the automated Abbott Alinity C analyzer (AFCAL).
  • To assess the clinical utility and efficiency of AFCAL in a core laboratory setting.

Main Methods:

  • Imprecision studies (5-day) evaluated assay variability.
  • Reportable range determination was performed.
  • Deming regression and Bland-Altman analysis compared AFCAL to predicate ELISA methods.
  • Retrospective clinical performance was assessed in 62 IBD patients.

Main Results:

  • Excellent imprecision: within-run 5.3%, between-day 2.5%.
  • Reportable range: 30-20,000 µg/g.
  • Strong correlation (r=0.99) and low bias (1.8%) with predicate ELISA.
  • AFCAL implementation improved turnaround time, reduced costs, and increased clinician satisfaction.

Conclusions:

  • The automated AFCAL assay is a reliable and efficient method for fecal calprotectin measurement.
  • AFCAL offers significant advantages over traditional ELISA methods for IBD management.
  • This validation supports the integration of AFCAL into routine clinical laboratory testing for IBD patients.