Frequency of Seroconversion in Aquaporin-4 Antibody Testing: Insights From Real-World Data

Tatchaporn Ongphichetmetha1,2, Mengke Du1, Nisa Vorasoot3,4

  • 1Neurology, Cleveland Clinic Neurological Institute, Mellen Center for Multiple Sclerosis, Cleveland, Ohio, USA.

Insights

Repeat testing for aquaporin-4-immunoglobulin G (AQP4-IgG) is rarely beneficial after a negative cell-based assay (CBA). Seroconversion is uncommon after negative immunofluorescence assays (IFA), suggesting limited yield from serial AQP4-IgG testing.

Area of Science:

  • Neurology
  • Immunology
  • Clinical Diagnostics

Background:

  • Aquaporin-4-immunoglobulin G (AQP4-IgG) antibodies are key biomarkers for neuromyelitis optica spectrum disorder (NMOSD).
  • Clinicians frequently retest AQP4-IgG to detect seroconversion, especially when initial results are negative.
  • Cell-based assays (CBA) offer superior sensitivity compared to older immunofluorescence assays (IFA).

Purpose of the Study:

  • To evaluate the frequency of AQP4-IgG seroconversion in patients with an initial negative test result.
  • To compare seroconversion rates between patients tested with CBA versus IFA.

Main Methods:

  • Retrospective cohort study of 451 patients tested between 2006 and 2024 at Cleveland Clinic.
  • Analysis of serial AQP4-IgG testing results based on the initial assay method (CBA or IFA).

Main Results:

  • No seroconversion was observed in 263 patients initially tested by CBA.
  • Seroconversion occurred in 4.3% (7/170) of patients initially tested by IFA.
  • Repeat AQP4-IgG testing showed a low yield for seroconversion, particularly after negative CBA results.

Conclusions:

  • Negative CBA results for AQP4-IgG are highly reliable, with minimal risk of subsequent seroconversion.
  • Repeat AQP4-IgG testing is generally not recommended after a negative CBA.
  • Serial testing may have limited utility after a negative IFA, but is less indicated than previously thought.

Related Concept Videos