Sequential versus Simultaneous Quantitative Analysis of Biomarkers in Individual Cells by ICP-MS and Mass Cytometry:

Ángela de la Rosa-Díaz1,2, Christian Sordo-Bahamonde2,3,4, Segundo Gonzalez2,3,4

  • 1Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, c/Julián Clavería 8, Oviedo 33006, Spain.

Analytical Chemistry
|November 12, 2025
PubMed

Insights

Accurate elemental quantification in single cells using inductively coupled plasma mass spectrometry (ICP-MS) is challenging. This study optimizes ICP-MS methods for comparable results with mass cytometry (CyTOF), validating its use for immunotherapeutic agent quantification.

Area of Science:

  • Analytical Chemistry
  • Cell Biology
  • Biotechnology

Background:

  • Accurate elemental quantification in single cells is vital for biological research.
  • Inductively coupled plasma mass spectrometry (ICP-MS) with nonsimultaneous mass analyzers presents technical challenges for elemental quantification.
  • Optimizing quantification strategies is essential for reliable single-cell analysis.

Purpose of the Study:

  • To establish conditions for comparable sequential and simultaneous elemental detection in single cells using ICP-MS.
  • To compare single-cell ICP-MS (SC-ICP-MS) with mass cytometry (CyTOF) for quantifying immunotherapeutic agent-target interactions.
  • To validate a unified calibration strategy for both SC-ICP-MS and CyTOF.

Main Methods:

  • Sequential and simultaneous elemental quantification using ICP-MS.
  • Optimization of cell storage and dilution conditions for ICP-MS analysis.
  • Comparison of SC-ICP-MS and CyTOF for quantifying Rituximab-CD20 interactions on MEC-1 cells labeled with Lutetium (Lu).

Main Results:

  • Cell storage and dilution conditions were found to be more critical than cell deposition for ICP-MS quantification.
  • SC-ICP-MS and CyTOF yielded comparable results for CD20 expression levels (approx. 1 × 10^4 receptors/cell).
  • A slight bias towards higher results was observed with SC-ICP-MS compared to CyTOF, potentially due to data treatment differences.

Conclusions:

  • Optimized ICP-MS protocols enable reliable elemental quantification in single cells.
  • SC-ICP-MS and CyTOF are suitable for quantifying immunotherapeutic agent binding with comparable results.
  • The study validates a unified calibration strategy for both SC-ICP-MS and CyTOF, applicable to clinical samples.

Related Concept Videos

Mass Spectrometry: Complex Analysis01:21

Mass Spectrometry: Complex Analysis

Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
2.1K
Tandem Mass Spectrometry01:21

Tandem Mass Spectrometry

Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and reduce chemical noise during analyte detection. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called...
3.0K
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview01:19

Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
2.7K