Video Experimental Relacionado
Updated: Jan 8, 2026

Modified Octopus Technique for Thoracoabdominal Aortic Aneurysm
Published on: August 1, 2025
Comparación de endoinjertos modificados por el médico Zenith de Cook y Relay Plus para aneurismas aórticos complejos
Samuel Saers1, Wajdi Alrawi2, Emiel W M Huistra3
1Department of Thoracic and Vascular Surgery, Linköping University Hospital, Linköping, Sweden; Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.
Background:
Complex abdominal aortic aneurysms (CAAA) or thoracoabdominal aneurysms (TAA) often require urgent repair, but custom-made devices are unavailable in such settings. Physician-modified endografts (PMEGs) allow for rapid, patient-specific fenestrated or branched endografts. This study assessed clinical outcomes of CAAA and TAA PMEGs performed at a single Swedish center and compared results using the Cook Zenith and Relay Plus platforms.
Methods:
A retrospective single-center review was conducted on all CAAA and TAA PMEG procedures at our center from January 2012 to December 2023. Data were collected retrospectively. Primary outcomes included mortality and technical success; secondary outcomes were major adverse events (MAEs) within 30 days, reinterventions, and bridging stentgraft patency.
Results:
All patients were treated using either the Zenith or Relay Plus systems. 42 patients were included (20 Zenith, 22 Relay Plus). Relay Plus was used more frequently in TAA (0% vs 36%, P = .004). Median follow-up duration was 4.8 years (1.8-8.2 years). Technical success was 90% in both groups. Four-year all-cause survival was 64% in both subgroups (p = .49). MAEs within 30 days occurred in 12% (N = 5) without significant subgroup differences. Freedom from reintervention was 69% in the Zenith group, and 59% in the Relay Plus group at 4 years (p = .34). Bridging stentgraft primary, primary assisted and secondary patencies were 91%, 97%, and 98%, respectively at four years.
Conclusion:
PMEGs using both platforms offer favorable outcomes in mortality, technical success, MAEs, and target vessel patency. Despite Relay Plus being used in more complex anatomies, no significant outcome differences were observed. This study supports the use of both Zenith and Relay Plus in the construction of PMEGs.