Evidencia preclínica mejorada de ingestión completa de coágulos con la bomba inteligente ALGO que respalda la
Amer Mitchelle1,2, Pervinder Bhogal3, Ravi Gandhi4
1The MIRI Centre, Ingham Institute for Applied Medical Research, Liverpool, Australia. amer@nhs.net.au.
Purpose:
To compare the ALGO Smart Pump (ALGO), operating in Adaptive Pulsatile Aspiration (APA™) and Static modes, with the Penumbra ENGINE during in-vitro mechanical thrombectomy, and to determine whether adaptive pulsatile aspiration improves complete clot ingestion (CCI) and reduces aspiration time across catheters sizes and catheter-to-artery (C/A) ratios.
Methods:
A pre-specified integrative analysis of two previously completed in-vitro thrombectomy studies was conducted using a standardized middle cerebral artery occlusion model. A total of 720 thrombectomy procedures were performed by two experienced operators using twelve commercial catheters (six small-bore, six large-bore). Three aspiration modalities were tested: ALGO APA™, ALGO Static, and Penumbra ENGINE continuous aspiration (n = 30 per catheter-pump pairing). The primary endpoint was CCI; secondary endpoint was aspiration time. Outcome analysis included two-way ANOVA and Chi-squared tests, with catheter, operator, and C/A ratio terms.
Results:
The ALGO demonstrated a higher median CCI rate than Penumbra (81.4% vs. 56.1%, χ2(1) = 52.364, p < 0.001). Aspiration times were significantly shorter with ALGO, particularly in APA™ mode (43.8 s vs. 60.6 s, p < 0.001). Two-way ANOVA demonstrated significant interaction between catheter size and pump type (p < 0.001), with ALGO notably improving small-bore catheter performance. CCI correlated positively with catheter-to-artery (C/A) ratio across both systems, with ALGO maintaining higher success across all ratios.
Conclusion:
Adaptive pulsatile aspiration enhances clot ingestion efficiency and reduces aspiration time in an in-vitro thrombectomy model, particularly for smaller catheters. These findings support further translational studies on optimizing aspiration dynamics to enhance first-pass success in endovascular stroke therapy.