自宅聴音検査と院内尿流測定の相関評価:新しいゴールドスタンダードに向けて
Raeven S Grant1, Kymora B Scotland1, Karan N Thaker1
1Department of Urology, David Geffen School of Medicine, University of California-Los Angeles, Los Angeles, CA.
Objectives:
To determine whether audio-based uroflowmetry is equivalent in accuracy and reliability to the in-clinic gold standard.
Methods:
Men over age 18 with a chief complaint of LUTS conducted a standard in-office uroflowmetry and used the Emano Flow phone application for one week to conduct audio-based uroflowmetry. Pearson correlation coefficient (PCC) was utilized to assess the relationship between conventional and the mean of audio-based Qmax after adjusting for the effect of void volume by performing residualization.
Results:
Thirty-two participants were recruited. On average, a patient's home volume was 2.18x (SD 1.67) higher than their clinic volume. Prior to adjusting for void volume, conventional and the mean of multiple at-home maximum flow rates were not correlated (PCC=0.21, p=0.31). After adjusting for void volume, conventional and mean of multiple at-home Qmax were significantly correlated (PCC=0.41, p = 0.04).
Conclusion:
Prior to adjusting for void volume, conventional and at-home maximum flow rates were not correlated. This can be explained, in part, by patients having one uncharacteristically low in-clinic on-demand void. Once we adjusted for void volume, office and at-home Qmax were correlated. As office Qmax may not accurately capture a patient's natural voiding patterns, our findings underscore the potential value of at-home testing. Further work is needed to improve the tools that we use for diagnosing and evaluating LUTS.
関連する概念動画
Urodynamic Studies: Uroflowmetry
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urine Studies I: Urinalysis
Urine Studies II: Urine Culture and Sensitivity Test
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Nursing Assessment of the Genitourinary System III: Percussion and Auscultation