関連する実験動画
Updated: Jan 8, 2026

Implantation of the Syncardia Total Artificial Heart
Published on: July 18, 2014
心臓肉腫に対する心臓全摘除後の体外循環構成におけるHeartMate 6
Hrvoje Gasparovic1, Maja Cikes2, Kristina Krželj1
1University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia.
Abstract:
We present the case of a 34-year-old patient with an unresectable cardiac sarcoma who underwent a total cardiectomy followed by implantation of 2 HeartMate 3 devices in a total artificial heart configuration. Surgical treatment of cardiac sarcomas ranges from palliative debulking to heart transplantation. In contrast to conventional total artificial heart placement, complete removal of both atrioventricular connections was mandated by the underlying pathologic process. Hemodynamic performance of 2 continuous flow pumps in the absence of native atria and the heart reservoir function depends on balancing preload, afterload, and individual pump rotations. The unreliability of conventional monitoring parameters in this clinical scenario makes hemodynamic management challenging.