冠動脈造影CTおよびOCTデータの包括的な融合ワークフローと3D印刷検証:経皮的冠動脈形成術デバイスのための患者固有のテスト環境の進歩
Marc Ilic1, Jonas Häner1, Julian Lehmann1
1Department of Cardiology, University Hospital Bern, Freiburgstrasse 20, CH-3010, Bern, Bern, Switzerland.
Objective:
To create high-resolution, patient-specific 3D coronary artery models aimed at developing digital models and functional phantoms for the testing of cardiac catheterization devices.
Methods:
Using coronary computed tomography angiography (CCTA) and optical coherence tomography (OCT), coronary artery lesions were identified and quantified. Imaging data were fused using a custom-made workflow to create highly accurate digital 3D models. For validation of the workflow, coronary artery phantoms were fabricated using additive manufacturing. An OCT was then conducted on the 3D printed phantom, and the developed workflow was applied to generate a derivative model, which was subsequently compared to the original.
Results:
CCTA and OCT datasets from 15 patients were successfully collected and used to develop patient-specific 3D coronary artery models, including detailed inner shells, calcifications, outer wall structures, and side branches. Of these, 13 out of 15 3D printed phantoms were successfully validated and compared to their corresponding original model. The median vertex deviation of the derivative model was 0.15 (0.140.17) mm. The median absolute stenosis difference between the derivative model and the original model was 3 (1-5)%AS.
Conclusion:
We present a novel workflow to produce high-resolution patient-specific phantoms of coronary arteries.