家庭用血圧計の2機種を家庭的環境で比較検証した結果、血圧値に顕著な差と精度低下が認められた
Joachim Zahnd1, Brian Thompson1, Pierre-Antonin Rigon1
1Service of nephrology and hypertension, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
Abstract:
Home blood pressure monitoring (HBPM) is essential for long-term hypertension management, but its accuracy and reliability is questionable due to user inconsistencies and non-standard usage conditions. This study compared two validated HBPM devices - a wrist-based and a reference upper-arm monitor - in a home-like setup where participants placed the devices themselves. A total of 121 participants underwent four concurrent blood pressure (BP) measurements, with two taken in each arm, followed by a crossover of devices between arms after the initial two readings. Wrist-derived BP readings were higher than upper-arm measurements. With higher blood pressure levels, both devices exhibited greater bias, accompanied by reduced precision in systolic BP measurements. Additionally, hypertension classification showed only moderate agreement (Cohen's kappa=0.58). The wrist monitor tended to over-diagnose hypertension and exhibited greater variability than the upper-arm device. These findings highlight the need for more guidance and education as well as critical evaluation of home blood pressure measurements provided by patients.
関連する概念動画
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)


