免疫グロブリンおよび補体タンパク質の定量のための免疫混濁測定法における酸化的干渉
Bruna Garlet Rossato1, Rafael Noal Moresco1
1Laboratory of Clinical Biochemistry, Department of Clinical and Toxicological Analysis, Center of Health Sciences, Federal University of Santa Maria, Santa Maria, RS, Brazil; Graduate Program in Pharmaceutical Sciences, Center of Health Sciences, Federal University of Santa Maria, Santa Maria, RS, Brazil.
Abstract:
Immunoturbidimetric assays are widely used to quantify immunoglobulins and complement proteins but are susceptible to interference. This study evaluated the effects of oxidative agents, including the Fenton reaction and hypochlorous acid (HOCl), on these assays. Serum pools were oxidized in vitro, and IgA, IgG, IgM, C3, and C4 were measured. Oxidative conditions increased measured concentrations of IgA, IgM, C3, and C4, whereas IgG was unaffected by HOCl. Increased oxidative stress index values and reduced thiol levels confirmed oxidative modification, indicating potential impairment of assay accuracy under oxidative stress.