2種類の延長焦点眼内レンズ(SymfonyおよびVivity)の視覚および主観的結果の比較:薄明かり下での性能に焦点を当てて
Joaquin O De Rojas1, Helga P Sandoval2, Richard Potvin3
1Center For Sight/US Eye, Sarasota, FL, USA.
Purpose:
To evaluate outcomes after implantation of two extended depth of focus IOLs, with emphasis on performance in dim light.
Setting:
Two clinical practices in the USA.
Design:
Prospective, double masked, randomized clinical trial.
Methods:
Subjects were randomized to receive Symfony Optiblue or Vivity IOLs. Objective optical quality, aberrations, and simulated visual disturbances were measured before and 3 months after surgery. Binocular visual acuity (VA), the defocus curve and low contrast VA were measured under mesopic lighting conditions. Patient questionnaires were administered and undirected patient comments were recorded.
Results:
Thirty-one subjects in each group completed the study. Refractive outcomes were not significantly different. The Symfony lens provided about a half line better monocular mesopic VA at distance when uncorrected and best corrected (p < 0.05). There were no differences in binocular visual function between lenses except for the mesopic defocus curve at -1.00 D and -1.50 D, with the Symfony lens providing better VA at those vergences (p < 0.02). Objective optical measures improved after surgery, with improvements not different by group. There were no differences in the size or intensity of simulated glare, but simulated halos and starburst size and intensity were significantly worse with the Symfony lens (p < 0.05). Satisfaction with distance and intermediate vision in both groups was high. Reported visual disturbances were not significantly different between groups.
Conclusions:
The two lenses provided similar visual performance in dim light, with the Symfony lens showing slightly better VA, but with more reports of halos and starbursts.
関連する概念動画
Predicting Reaction Outcomes
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Uniform Depth Channel Flow
Depth Perception and Spatial Vision
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...