追加吸引チャンネルによる緊急内視鏡検査効率の向上:invitro評価
André Sasse1, Thomas Roland Heiduk1, Marvin Scheunemann2
1Clinic for Gastroenterology, Gastrointestinal Oncology and Endocrinology, University Medical Center Göttingen, Göttingen, Germany.
Background And Study Aims:
Emergency endoscopic interventions for upper gastrointestinal hemorrhage are frequently hampered by presence of blood clots and food debris. This study aimed to assess whether integration of an additional suction channel (ASC) onto a standard gastroscope enhances efficiency of aspirating clots and viscous fluids.
Patients And Methods:
A 5.3-mm suction catheter was used as an ASC mounted on a 2.8-mm standard gastroscope. Suction efficacy was evaluated using gastroscopes with working channel diameters of 2.8 mm, 3.7 mm, and 6 mm in vitro. Defined volumes of water, fruit yogurt, and coagulated blood were aspirated, and time required for complete evacuation was measured. Each setup was tested with and without the BioVac system.
Results:
The ASC significantly enhanced suction performance across all test media. Notably, the 2.8-mm gastroscope with ASC outperformed all other configurations in aspirating water and yogurt. For clotted blood, the ASC significantly improved evacuation times compared with all other setups besides 6-mm + BioVac.
Conclusions:
A standard gastroscope equipped with an ASC significantly enhances suction performance in an in vitro model, outperforming gastroscopes with larger working channels. These findings warrant further validation in an ex vivo model to determine their clinical applicability.
さらに関連する動画
08:53Author Spotlight: Advancing Erythropoiesis Research - A Simplified Pipeline for Assessing Hematopoietic Stem Cell Function in Myelodysplastic Syndromes
Published on: January 10, 2025
08:06Enhancement Method of Surface Acoustic Wave-Atomizer Efficiency for Olfactory Display
Published on: November 14, 2018
関連する概念動画
Conjugate Addition (1,4-Addition) vs Direct Addition (1,2-Addition)
Conjugate addition results in a thermodynamically stable product. The reaction retains the stronger C=O bond at the expense of the weaker C=C π bond. The process is slow as the β carbon is less electrophilic than the carbonyl carbon.
Direct addition products are...
Tracheostomy Suctioning II: Procedure
Emerging Adulthood
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Conjugate Addition of Enolates: Michael Addition