ファントムを用いた超解像および通常解像設定における2つの商用深層学習CT再構成アルゴリズムの性能比較
Joël Greffier1, Catherine Roy2, Djamel Dabli3
1IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, Nîmes, France. joel.greffier@chu-nimes.fr.
Objective:
We compared a super-resolution deep learning image reconstruction (SR-DLR) algorithm with a normal-resolution (NR)-DLR algorithm according to radiation dose for abdominal computed tomography (CT).
Materials And Methods:
An image-quality phantom was scanned with an energy-integrating detectors CT unit at three volume CT dose index radiation dose levels (12.7, 5.9, and 3 mGy). Images were reconstructed using a 1,0242 matrix for SR-DLR and a 5122 matrix for NR-DLR, for three DLR levels (level-1, level-2, and level-3). Noise power spectrum (NPS) and task-based transfer function (TTF) for iodine and Solid Water® inserts were computed; TTF values at 50% (f50, mm-1) were used to quantify spatial resolution. The detectability index (d') was computed for two simulated lesions.
Results:
Noise magnitude values were lower with SR-DLR than with NR-DLR for level-2 (-27.6 ± 3.8%) and level-3 (-43.5 ± 1.4%), the opposite for level-1. Average NPS spatial frequency was higher with SR-DLR than with NR-DLR for all radiation dose levels for level-1 (55.9 ± 16.7%) and level-2 (20.1 ± 13.9%) and the opposite for level-3, except at 12.7 mGy. For both inserts, f50 was higher with SR-DLR than with NR-DLR at each radiation dose and DLR level. For simulated lesions and all DLR levels, d' values were higher with SR-DLR than with NR-DLR (level-1, 6.0 ± 2.0%; level-2, 45.7 ± 5.0%; level-3, 75.2 ± 7.3%).
Conclusion:
Compared to NR-DLR, SR-DLR improved spatial resolution and detectability of simulated abdominal lesions; image noise was reduced with SR-DLR only for level-2 and level-3, while image texture was better for level-1 and level-2.
Relevance Statement:
Super-resolution DLR with a 1,0242 matrix size improved spatial resolution and detectability of simulated abdominal lesions compared to normal-resolution DLR. Validation in clinical settings is necessary before translation into routine practice.
Key Points:
The performance of a new deep learning super-resolution image reconstruction algorithm (SR-DLR) was compared to a normal-resolution (NR)-DLR algorithm using an image-quality phantom for an abdominal energy-integrating detector CT protocol. SR-DLR with a 1,0242 matrix improved spatial resolution and detectability of simulated abdominal lesions compared to NR-DLR with a 5122 matrix. Using SR-DLR, therefore, presents numerous prospects for improving abdominal CT images and a high potential for reducing the radiation doses.
さらに関連する動画
関連する概念動画
Super-resolution Fluorescence Microscopy
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Trial and Error and Algorithm
Normal Stress
When a rod is under axial loading, the internal forces and corresponding stress are normal to the plane of the section, so it is termed normal stress. It's important to...
Normal Distribution
Variation: Normal Distribution, Range, and Standard Deviation