新生児高ビリルビン血症における測定精度を向上させた経皮的ビリルビン測定器の改良
Kosuke Koyano1, Makoto Arioka2, Yuta Noguchi3
1Faculty of Medicine.
Background:
Conventional transcutaneous bilirubinometers often underestimate total bilirubin (TB) levels at high concentrations, limiting their clinical utility in cases of neonatal hyperbilirubinemia. We aimed to address this shortcoming by developing and testing a novel type of advanced transcutaneous bilirubinometer that may enhance the early identification of severe hyperbilirubinemia, potentially reducing the need for invasive testing and exchange transfusions.
Methods:
We developed a novel transcutaneous bilirubinometer with adjustable light intensity and detection distances. Transcutaneous bilirubin (TcB) measurements were obtained using both the novel device and a conventional JM-105 bilirubinometer in 66 instances in 62 neonates (≥36 weeks gestation). TB values measured in patient blood samples were used as reference values for calculating correlations and error metrics (i.e., mean squared error [MSE] and mean absolute error [MAE]).
Results:
The novel device showed strong correlation with TB values across all concentrations (R² = 0.96), including those ≥15 mg/dL (R² = 0.84), outperforming the JM-105 device (R² = 0.60 for TBs ≥15 mg/dL). The novel device also yielded lower MSE (1.53 vs. 2.95) and MAE values (1.00 vs. 1.39) than the JM-105.
Conclusions:
Our novel transcutaneous bilirubinometer demonstrated improved accuracy at higher bilirubin concentrations compared with a conventional JM-105 device.
関連する概念動画
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...