血液製剤施設におけるヘモグロビン測定不確かさ推定における新たな実践
Maria Helena Sousa Timóteo1, Ana Sofia R Tavares1, Paulo Pereira1
1Lisbon School of Health, Polytechnic University of Lisbon, Portugal.
Abstract:
Measurement uncertainty (MU) is increasingly recognized as a critical analytical parameter within ISO and European regulatory frameworks. In transfusion laboratories, where hemoglobin testing determines donor eligibility and informs transfusion decisions, MU directly impacts patient and donor safety. This study applies ISO/TS 20914:2019 to estimate MU in hemoglobin testing using the cyanmethemoglobin method in EDTA-K3 venous blood. A parametric simulation model informed by EFLM biological variation data was used to evaluate three scenarios: (1) within-laboratory imprecision, (2) addition of calibrator uncertainty, and (3) incorporation of significant bias uncertainty. All scenarios met desirable (2.7 %) and minimum (4.1 %) performance targets for allowable uncertainty. These findings reinforce the integration of MU estimation as a European and ISO-endorsed best practice in transfusion laboratory medicine. This approach supports traceable, risk-informed decision-making, aligning transfusion services with ISO 15189:2022 requirements.