ZW-30 SS-OCT生体測定装置とMS-39 SD-OCT/プレシドトポグラファーの前眼部測定における一致性
Chak Seng Lei1,2, Qinghao Meng3, Qingyi Gao1,2
1Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University; NHC Key Laboratory of Myopia (Fudan University), Key Laboratory of Myopia, Chinese Academy of Medical Sciences, Shanghai, China.
Purpose:
To evaluate the agreement of anterior segment biometric measurements between a new swept-source optical coherence tomography (SS-OCT) biometer (ZW-30) and a spectral-domain optical coherence tomography combined with Placido-disk topography (SD-OCT/Placido topographer, MS-39).
Setting:
Eye and ENT Hospital of Fudan University, Shanghai, China.
Design:
Prospective cross-sectional study.
Methods:
This prospective, cross-sectional study recruited 70 healthy eyes from 70 participants. Each eye underwent three consecutive scans with both the ZW-30 and MS-39. Parameters analyzed included central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), J0, J45 astigmatism vector components and corneal diameter (CD). Inter-device differences were evaluated using paired t-tests, Bland-Altman analysis with 95% limits of agreement (LoA) assessed the agreement.
Results:
Significant differences were observed in all parameters measured, except for J0. Bland-Altman analysis indicated high agreement for CCT (mean difference: 5.38 ± 3.67 μm; 95% LoA: -2.73 to 13.48 μm), ACD (-0.11 ± 0.03 mm; -0.18 to -0.03 mm), LT (-0.06 ± 0.05 mm; -0.17 to 0.05 mm), Km (0.07 ± 0.16 D; -0.33 to 0.47 D), J0 (0.02 ± 0.12 D; -0.31 to 0.27 D), J45 (-0.08 ± 0.09 D; -0.32 to 0.16 D). A larger discrepancy was observed for CD (0.48 ± 0.13 mm; 0.17 to 0.79 mm).
Conclusions:
The ZW-30 SS-OCT biometer and the MS-39 SD-OCT/Placido topographer demonstrated high agreement for most anterior segment parameters, indicating their interchangeability in clinical practice. However, caution is advised when using CD measurements interchangeably due to the observed systematic differences between devices.
関連する概念動画
Topographic Surveying and Contours
Plotting of Topographic Maps
Muscles of the Anterior Neck
Crystal Field Theory - Octahedral Complexes
To explain the observed behavior of transition metal complexes (such as colors), a model involving electrostatic interactions between the electrons from the ligands and the electrons in the unhybridized d orbitals of the central metal atom has been developed. This electrostatic model is crystal field theory (CFT). It helps to understand, interpret, and predict the colors, magnetic behavior, and some structures of coordination compounds of transition metals.
CFT focuses on...
Uncertainty in Measurement: Reading Instruments
Measures of Central Tendency