概括
带有植入式心脏变压器 (ICD) 的患者可以安全地穿过电子物品监控系统. 然而,在这些系统附近停留可能会导致不适当的ICD冲击,对患者的安全构成风险.
科学领域:
- 心脏病学 心脏病学
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
背景情况:
- 植入式心脏变压器 (ICD) 的不适当冲击与电子物品监控 (EAS) 系统有关.
- 当患者通过EAS系统或停留在EAS系统附近时,与ICD相关的特定风险需要进一步澄清.
研究的目的:
- 评估ICD功能的安全性,当患者通过或暴露于EAS系统时.
- 调查ICD和EAS系统之间可能导致不良事件的潜在相互作用.
主要方法:
- 170名患有ICD的受试者在短时间的介绍和长期暴露 (6英寸内2分钟) 期间对三种常见的EAS系统进行了评估.
- 在这些暴露期间,ICD功能被监测,包括ICD节奏的场景.
主要成果:
- 在对EAS系统进行10至15秒的步行过程中,没有观察到对ICD功能的负面影响.
- 在长期暴露期间,169名受试者中有19人发生相互作用;其中7人具有临床意义,有可能产生不适当的ICD冲击.
- 12名受试者经历了轻微的相互作用.
结论:
- 通过EAS系统走路对于ICD患者来说是安全的.
- 长时间暴露或在EAS系统附近逗留可能会引发不适当的ICD冲击.
更多相关视频
相关概念视频
Pulse rhythm
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Instrumentation Amplifier
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Instrument transformers, comprising voltage transformers (VTs) and current transformers (CTs), play crucial roles in power substations by providing isolated replicas of current or voltage for measurement and protection purposes. Voltage transformers reduce the primary voltage to levels suitable for relay operation and measurement, while current transformers scale down the primary current. The primary winding of a current transformer often consists of a single turn, achieved by threading the...
Differential Relays
Differential relays are used to protect generators, buses, and transformers by comparing electrical quantities at different points. When a fault occurs, the difference in current between the two points triggers the relay to operate, opening the circuit breaker. Under normal conditions, the current entering (i1) and leaving (i2) a generator are equal. When a fault occurs, however, these currents become unequal, and the difference current flows in the relay operating coil, causing the relay to...
Holter Monitor: 24-Hour Monitoring
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...