在一个新的胆叶叶片机械中,叶片泄漏:TRI技术
Tomaso Bottio1, Dino Casarotto, Gaetano Thiene
1Department of Cardiovascular Surgery, University of Padua Medical School, Via A. Gabelli, 61, 35121 Padova, Italy.
Circulation
|May 7, 2003
概括
由于传单逃逸而导致的机械心脏膜衰竭发生在两名接受TRI Technologies假肢的患者身上. 这种结构故障是由于传单骨折造成的,导致植入程序停止.
科学领域:
- 心血管外科心血管外科
- 生物医学工程 生物医学工程
- 材料科学 材料科学 材料科学
背景情况:
- 传单逃逸是机械心脏门结构性故障的关键模式.
- 以前的报道记录了各种机械门设计中的这种并发症.
- 这项研究详细介绍了TRI Technologies门假肢的两起传单泄漏案例.
研究的目的:
- 报告和分析TRI Technologies门假肢患者的两起传单泄漏事件.
- 为了调查这些特定设备中结构门故障的原因.
- 评估对TRI Technologies门假肢的持续使用的影响.
主要方法:
- 对接受TRI Technologies门假肢的患者进行了回顾性分析.
- 临床数据审查,包括患者人口统计,手术程序和结果.
- 检查已扩展的假肢,以确定故障的机制.
主要成果:
- 在2000年11月至2002年2月期间,TRI Technologies在34名患者身上植入了36个门假肢.
- 两名患者经历了灾难性的假肢传单逃逸,原因是传单旋转系统骨折.
- 一名患者在术后死亡,而另一名患者需要紧急重新手术.
结论:
- 观察到的结构门故障引起了对TRI Technologies门假肢的耐用性严重的担忧.
- 由于高失败风险,导致植入程序中断.
- 需要对这些假肢的设计和制造进行进一步的研究.
相关概念视频
Pivot Bearings
2.8K
In mechanical systems, bearings are crucial in facilitating relative motion between two components while minimizing friction and wear. They help distribute various loads (radial, axial or a combination of both loads) across machinery parts, ensuring smooth and efficient operation.
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...
2.8K
Heart Valves
14.9K
The human heart is a complex organ with an intricate system of valves that regulate blood flow. There are two main types of valves: atrioventricular (AV) valves and semilunar valves.
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
14.9K
Mechanical Systems
901
Mechanical systems are analogous to to electrical networks where springs and masses play similar roles to inductors and capacitors, respectively. A viscous damper in mechanical systems functions similarly to a resistor in electrical networks, dissipating energy. The forces acting on a mass in such systems include an applied force in the direction of motion, counteracted by forces from the spring, a viscous damper, and the mass's acceleration. This interplay of forces is mathematically...
901
Mitral Regurgitation III: Medical Management
686
Mitral regurgitation (MR) is characterized by retrograde blood circulation from the left ventricle into the left atrium due to inadequate mitral valve closure. The severity of the condition, symptoms, and underlying cause determine treatment strategies.Monitoring and Pharmacological TreatmentPatients with mild to moderate MR typically do not need immediate intervention but regular monitoring to assess progression and guide treatment. Patients with mild MR should have an echocardiogram every 3-5...
686
Mitral Stenosis III: Medical Management
535
Mitral stenosis, a condition marked by the narrowing of the mitral valve, necessitates an integrated approach for effective management. This approach includes preventative measures, medical therapy, and surgical interventions to reduce symptoms and prevent complications.PreventionPrevention of mitral stenosis primarily focuses on reducing the incidence of bacterial infections, particularly streptococcal infections, which can lead to rheumatic fever and subsequent valvular damage. Timely...
535
Bioreactor Controls-I
94
Maintaining optimal conditions within fermenters is essential for maximizing microbial productivity and ensuring process efficiency. This lesson focuses on key parameters—temperature, foam, pH, carbon dioxide, oxygen, and pressure—and their precise measurement and control strategies in fermentation systems.Temperature ControlTemperature regulation is critical due to the exothermic nature of many fermentation processes. In small laboratory fermenters, temperature is commonly...
94