与光学低一致性反射生物计和SS-OCT生物计相比,一个联合扫描源光学一致性和反射生物计的折射预测性和生物计协议
Bjørn Gjerdrum1, Kjell Gunnar Gundersen1, Christian Nilsen1,2
1Ifocus Eye Clinic, Stavanger, Norway.
Clinical ophthalmology (Auckland, N.Z.)
|May 30, 2023
概括
这项研究比较了白内障手术的三种生物计,发现所有生物计都显示出良好的折射可预测性. 阿戈斯扫描源OCT生物仪显示了最低的折射预测误差,尽管差异在统计学上并不显著.
科学领域:
- 眼科医生 眼科 眼科
- 生物医学工程 生物医学工程
- 光学计量学 在光学计量学
背景情况:
- 准确的眼内镜 (IOL) 功率计算对于成功的白内障手术结果至关重要.
- 生物识别设备在获得精确的手术前测量方面发挥着关键作用.
- 对比不同的生物识别技术对于优化折射可预测性至关重要.
研究的目的:
- 为了评估一个细分轴长度 (AL) 计算扫源光学连贯性断层扫描 (SS-OCT) 生物仪和另一个SS-OCT生物仪,以及一个光学低连贯性反射计 (OLCR) 生物仪之间的折射可预测性协议.
- 描述折射结果,视力敏度,以及手术前生物识别参数的一致性.
主要方法:
- 在白内障手术后对129只眼睛进行了回顾性单臂研究.
- 使用两个SS-OCT设备 (Argos,Anterion) 和一个OLCR设备 (Lenstar 900) 收集的手术前生物识别数据.
- 用于IOL功率计算的Barrett Universal II公式;分析了折射预测误差 (RPE) 和绝对误差 (AE).
主要成果:
- 平均RPE:阿尔戈斯 (0.06 D),安特里昂 (-0.14 D),伦斯塔 (0.17 D).
- 阿尔戈斯的绝对RPE最低;伦斯塔的AE中位数最低,但差异没有统计学意义.
- 眼睛的百分比在±0.5 D RPE范围内:阿戈斯 (76%),安特里昂 (71%),伦斯塔 (78%). 在±0.5 D AE内的百分比:阿戈斯 (79%),安特里昂 (84%),伦斯塔 (82%). 没有显著的差异.
结论:
- 这三种生物计在白内障手术后都表现出良好的折射可预测性.
- 在设备之间的折射预测误差或绝对误差中没有发现统计学上显著的差异.
- 阿戈斯SS-OCT生物计显示了算术折射预测误差最低的数值.
相关概念视频
Imaging Biological Samples with Optical Microscopy
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Confocal Fluorescence Microscopy
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...