热处理和仪器的动态循环疲劳阻力在反向运动中
Thaís Kauana Magalhães Sobral1, Lucila Piasecki2, Flávia Sens Fagundes Tomazinho3
1Department of Dentistry, Centro Universitário UniDomBosco, Curitiba, Brazil.
European endodontic journal
|May 31, 2023
概括
与 Reciproc 和 X1 Blue 仪器相比,Pro-R 和 Reciproc Blue 仪器在动态循环疲劳测试中表现出优越的耐疲劳性. 这一发现对于内牙科仪器的选择至关重要.
科学领域:
- 牙周内科医院 牙周内科医院 牙周内科医院
- 材料科学 材料科学 材料科学
- 生物医学工程 生物医学工程
背景情况:
- 互换仪器在内牙治疗中被广泛使用.
- 了解它们的耐疲劳性对于临床成功和防止仪器分离至关重要.
- 热处理的变化可以显著影响仪器的性能.
研究的目的:
- 为了比较四种不同的热处理互旋内牙仪器的动态循环疲劳阻力.
- 在模拟的临床条件下评估热处理对这些仪器寿命的影响.
主要方法:
- 测试了48个反转仪器 (X1蓝色,Pro-R,反转,反转蓝色).
- 仪器在人工道 (60°角,5mm半径) 中经过循环疲劳,在37°C的石块中进行了循环疲劳.
- 记录了故障的时间和碎片长度,并进行了统计分析.
主要成果:
- 与X1 Blue和Reciproc仪器相比,Pro-R和Reciproc Blue仪器显示出明显更高的耐疲劳能力 (p<0.05).
- 互换仪表显示中等疲劳耐力,明显高于X1 Blue.
- 断层分析证实循环疲劳是所有测试仪器的故障模式.
结论:
- Pro-R 和 Reciproc Blue 仪器提供了对动态循环疲劳的增强抵抗力.
- 这些发现表明,特定的热处理可以提高互换性内牙科文件的耐用性.
- 临床选择具有优越耐疲劳的仪器可以降低手术并发症的风险.
相关概念视频
Fatigue
213
Fatigue occurs when materials rupture under repeated or fluctuating loads, even at stress levels far below their static breaking strength. It typically results in brittle failure, even for ductile materials. It is a critical consideration in designing machines and structural components subjected to repetitive or varying loads. The nature of these loadings can range from fluctuating loads like unbalanced pump impellers causing vibrations to repeatedly bending a thin steel rod wire back and forth...
213
Temperature Dependent Deformation
174
In a nonhomogeneous rod made up of steel and brass, restrained at both ends and subjected to a temperature change, several steps are involved in calculating the stress and compressive load. Due to the problem's static indeterminacy, one end support is disconnected, allowing the rod to experience the temperature change freely. Next, an unknown force is applied at the free end, triggering deformations in the rod's steel and brass portions. These deformations are then calculated and added...
174
Yield Criteria for Ductile Materials under Plane Stress
195
In designing structural elements and machine parts using ductile materials, it is crucial to ensure that these components withstand applied stresses without yielding. Yielding is initially determined through a tensile test, which evaluates the material's response to uniaxial stress. However, tensile stress is insufficient when components face biaxial or plane stress conditions This condition requires advanced criteria to predict failure.
The Maximum Shearing Stress Criterion, also known as...
The Maximum Shearing Stress Criterion, also known as...
195