用永久填充的Ge-phantom测量PET图像质量和量化的变化:一个多中心研究
O Sipilä1, J Liukkonen2, H-L Halme3
1HUS Diagnostic Center, Clinical Physiology and Nuclear Medicine, Helsinki University Hospital and University of Helsinki, P. O. Box 442, 00029, Helsinki, Finland. outi.sipila@hus.fi.
EJNMMI physics
|June 15, 2023
概括
这项研究评估了使用幻影的芬兰PET/CT扫描仪的变异性. 结果显示,定量准确度的变化很大,但大多数扫描仪通过优化满足认证标准.
科学领域:
- 核医学就是核医学.
- 医学成像物理 医学成像物理
背景情况:
- 芬兰临床PET [18F]FDG全身方案的定量准确性和图像质量 (IQ) 的评估变异性.
- 使用一个充满68Ge的NEMA/IEC智商幻影进行快照评估.
研究的目的:
- 评估临床上使用的PET [18F]FDG全身协议的量化和图像质量的变异性.
- 为了将恢复系数 (RCs) 与 EARL 18F 标准 2 认证 (EARL2) 限制进行比较.
- 为了研究图像噪声对定量参数的影响.
主要方法:
- 在两个供应商的14台PET-CT扫描仪上成像了一个NEMA/IEC智商幻影.
- 测量RCs (RCmax,RCmean,RCpeak) 的变化率,百分比背景变化率 (PBV),背景变化系数 (COV_BG) 和校正精度 (AOC).
- 使用临床和标准化协议进行了20次重复测量,并分析了平均图像 (AVI) 来研究图像噪声的影响.
主要成果:
- 常规方案显示RCmax的变化率高达68%,在排除的情况下降至36%. 扫描器内部的变化率为10%.
- 大多数RC都符合EARL2范围,但确切的极限在变化方面得到了满足. RCpeak 的性能比 RCmax 和 RCmean 的性能更强大.
- PBV,COV_BG和AOC各有所不同 (分别为2.3-11.8%,9.6-17.8%,4.8-32.0%). 平均图像降低了RC范围,PBV和COV_BG. 没有点传播函数 (PSF) 校正的协议显示AOC.较高.
结论:
- [18F]FDG协议的RC值的最大变化大约为60%.
- 适当的交叉校准的扫描仪与PSF校正通常满足EARL2 RC范围,但需要进一步优化精确的限制.
- RCpeak显示出对图像噪声的最强健性;与COV_BG不同的是,RC和PBV对噪声敏感.
更多相关视频
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
5.6K
10:02Whole-body PET/MRI of Pediatric Patients: The Details That Matter
Published on: December 19, 2017
14.6K
相关概念视频
Positron Emission Tomography
4.3K
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
4.3K
Imaging Studies II: Positron Emission Tomography and Scintigraphy
167
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
Fundamental Principles of PET
167
Radiological Investigation III: Pulmonary Angiogram and PET Scan
117
Radiological investigations are paramount in the diagnosis and management of various pulmonary diseases. Two essential investigations are the Pulmonary Angiogram and the Positron Emission Tomography (PET) Scan.
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
117