螺旋式中心线支架与直立支架放置对血液流速的影响
Yutaro Kohata1,2, Makoto Ohta2,3, Kazuyoshi Jin1,2
1Graduate School of Biomedical Engineering, Tohoku University, Sendai, Japan.
Frontiers in medical technology
|June 19, 2023
概括
与直立支架相比,螺旋式支架可以通过保持更高的速度保留来改善股骨支架的血液流动. 这项研究量化了由螺旋式支架放置所影响的流动动力学.
科学领域:
- 生物医学工程 生物医学工程
- 血管外科 血管外科
- 医疗成像医学成像
背景情况:
- 骨关节支架的目的是保持动脉通透性.
- 螺旋式支架配置已经显示出改善了临床通透性.
- 对螺旋式支架对血液流动的影响缺乏定量分析.
研究的目的:
- 量化估计流速并评估螺旋式支架放置的影响.
- 为了比较状与直立支架在股骨关节动脉中的血液动力学效应.
主要方法:
- 在三个健康的猪身上植入螺旋式和直立式支架.
- 使用血管图像的时间强度曲线 (TIC) 分析估计血液流速.
- 对动脉扩张和流动动力学的观察.
主要成果:
- 血管造影显示螺旋式支架的对比度较薄,这表明旅行速度更快.
- 与直立支架 (43.0%-68.0%) 相比,螺旋式支架植入保持了更高的速度保留率 (55.0%-71.3%).
- 动脉扩张因位置而异,在所有支架病例中都观察到.
结论:
- 螺旋式支架的放置可能会增强股骨动脉的流动动力学.
- 虽然螺旋式支架显示出更好的速度保留,但在本研究中,螺旋式支架和直立式支架之间没有发现统计学上显著的差异.
- 需要进一步的研究才能充分理解螺旋式支架设计的血液动力学优势.
相关概念视频
Bernoulli's Equation for Flow Normal to a Streamline
912
Bernoulli's equation for flow normal to a streamline explains how pressure varies across curved streamlines due to the outward centrifugal forces induced by the fluid's curvature. The pressure is higher on the inner side of the curve, near the center of curvature, and decreases outward to balance these centrifugal forces.
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines.
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines.
912
Steady, Laminar Flow in Circular Tubes
263
Hagen-Poiseuille flow describes a viscous fluid's steady, incompressible flow through a cylindrical tube with a constant radius R. This flow profile is often applied to understand fluid transport in narrow channels, such as capillaries. It serves as a foundational example of laminar flow. In this model, cylindrical coordinates (r,θ,z) are used to describe the radial (r), angular (θ), and axial (z) dimensions within the tube. For Hagen-Poiseuille flow, the velocity profile is...
263
Autoregulation of Blood Flow
2.4K
Autoregulation mechanisms are characterized by their inherent capacity for self-regulation without necessitating specific nervous stimulation or endocrine control. These mechanisms facilitate the adjustment of blood flow and, therefore, perfusion specific to each tissue region. This self-regulation encompasses chemical signals and myogenic controls.
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
2.4K
Blood Flow
70.4K
Blood is pumped by the heart into the aorta, the largest artery in the body, and then into increasingly smaller arteries, arterioles, and capillaries. The velocity of blood flow decreases with increased cross-sectional blood vessel area. As blood returns to the heart through venules and veins, its velocity increases. The movement of blood is encouraged by smooth muscle in the vessel walls, the movement of skeletal muscle surrounding the vessels, and one-way valves that prevent backflow.
70.4K
Steady, Laminar Flow Between Parallel Plates
246
Understanding steady, laminar flow between parallel plates is essential for analyzing and designing flow in narrow rectangular channels, commonly found in various water conveyance and drainage systems. The Navier-Stokes equations govern fluid motion and are generally challenging to solve due to their nonlinearity. However, simplifications are possible in certain cases, like the steady laminar flow between parallel plates. For this scenario, we assume steady, incompressible, laminar flow.
246
Streamlines, Streaklines, and Pathlines
1.4K
A streamline represents the trajectory that is always tangent to the fluid's velocity vector at any given point. The velocity of a fluid particle is always directed along the streamline, ensuring the particle continuously follows the streamline's path. Streamlines are particularly useful for visualizing the overall direction of flow in a fluid system, and they provide an instantaneous representation of the flow's velocity field. In steady flow, where conditions do not change over...
1.4K