概括
在强度炼期间,CORE可穿戴设备不准确地测量核心体温 (TC),无法检测冷却效应,并低估了炎热条件下的温度升高.
科学领域:
- 运动生理学 运动生理学
- 运动科学 运动科学 运动科学
- 生物医学工程 生物医学工程
背景情况:
- 核心体温 (TC) 监测对于节奏和预防耐力赛期间的热病至关重要.
- 可穿戴技术提供了实时TC反的潜力,有助于性能和安全.
研究的目的:
- 为了比较CORE可穿戴设备与胃肠道药丸的准确性,用于测量冷水摄入期间的TC和5公里跑步时间试验.
- 评估两种设备之间的TC测量差异的临床可接受性.
主要方法:
- 十二名参与者在摄入冷水后,在30°C的热量下完成了5公里的跑步时间试验.
- 核心体温同时使用CORE可穿戴设备和胃肠道药丸进行测量.
- 分析了设备之间的TC读数的差异,其值>±0.25°C被认为是临床上不可接受的.
主要成果:
- 与胃肠道药丸相比,CORE装置在计时试验之前和期间都显示了TC的显著低估.
- 核心装置无法检测冷水摄入的冷却效应.
- 在时间试验期间,CORE和GI药丸之间的TC的平均绝对差异为0.47±0.34°C.
结论:
- 核心可穿戴设备不能准确地反映核心体温的变化,在一个5公里运行时间试验在热.
- 该设备无法捕捉冷水的冷却效应,并低估了TC,这限制了其优化节奏和预防热病的实用性.
相关概念视频
Assessing Body Temperature - Rectal
4.9K
Rectal temperature measurement is considered the most precise method for assessing core body temperature and typically registers higher than oral temperature. For adults, the rectal thermometer should be inserted 1 to 1.5 inches into the rectum to obtain the most accurate reading.
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
4.9K
Assessing Body Temperature - Oral
780
Here are the steps to accurately measure oral temperature using an electronic thermometer:
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
780
Equipments Used to Measure Body Temperature
1.1K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
1.1K
Thermosensation
30.5K
Peripheral thermosensation is the perception of external temperature. A change in temperature (on the surface of the skin and other tissues) is detected by a family of temperature-sensitive ion channels called Transient Receptor Potential, or TRP, receptors. These receptors are located on free nerve endings. Those detecting cold temperatures are closer to the surface of the skin than the nerve endings detecting warmth. These thermoTRP channels, while temperature selective, have relatively...
30.5K
Assessing Body Temperature - Temporal Artery
597
Here is a stepwise guide to assessing the body temperature at the temporal artery using a temporal artery thermometer
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
597
Body Temperature
374
Body temperature reflects the equilibrium between heat production and heat loss within the body. Most heat is generated by metabolically active tissues, particularly the liver, heart, brain, kidneys, and endocrine organs. At rest, skeletal muscles contribute 20–30% of total heat production, but during vigorous exercise, this can increase up to 30–40 times.
The average body temperature is approximately 37°C (98.6°F) and typically ranges from 36.1–37.2°C...
The average body temperature is approximately 37°C (98.6°F) and typically ranges from 36.1–37.2°C...
374