对Atellica系统的精度和真实性验证研究
Alberto Vílchez Rodríguez1, Julia González Cantó1, Sara Esteve Poblador1
1Area of Biological Diagnosis, La Ribera University Hospital, Alzira, Spain.
Advances in laboratory medicine
|June 26, 2023
概括
这项研究验证了临床实验室分析仪的精度和真实性. 阿特利卡®解决方案系统符合制造商和实验室的不精确性规格,尽管可能需要进行校准调整.
科学领域:
- 临床化学 临床化学
- 实验室医学 实验室医学
- 分析化学 分析化学
背景情况:
- 验证的程序对于临床实验室至关重要.
- 精度和真实性是验证新测量方法的关键指标.
- 评估诊断仪器的分析性能对于可靠的患者结果至关重要.
研究的目的:
- 为了验证不同实验室分析仪进行的生化和免疫化学分析的精度和真实性.
- 为了比较Advia 1800,Immulite 2000,CentaurXP和Atellica®Solution系统的性能,请参阅下面的图片.
- 确保测量程序符合已确定的实验室和制造商规格.
主要方法:
- 使用Advia 1800,Immulite 2000,CentaurXP和Atellica®Solution对十种分析物的生物化学和免疫化学表征.
- 使用BioRad®和西门子®控制材料在五天内 (运行内和运行间) 进行不精确性评估.
- 使用血清样本通过差异分析和线性回归评估偏差.
主要成果:
- 对于所有分析品,运行内和运行间的不精确值低于制造商声称的值.
- 在偏差评估过程中,对某些分析品检测到一个比例的,不恒定的系统错误.
- 阿特利卡®解决方案显示变化系数验证了制造商和实验室的不精确规格.
结论:
- 阿特利卡®解决方案系统成功验证了不精确的规格.
- 校准条件需要对特定参数进行修订以优化性能.
- 建议在未来的验证研究中使用更广泛的样本范围.
相关概念视频
Uncertainty in Measurement: Accuracy and Precision
74.0K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.
74.0K
Accuracy, limits, and approximation
480
Accuracy, limits, and approximations are common in many fields, especially in engineering calculations. These concepts are imperative for ensuring that a given value is as close as possible to its true value.
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
480
Improving Translational Accuracy
11.7K
Base complementarity between the three base pairs of mRNA codon and the tRNA anticodon is not a failsafe mechanism. Inaccuracies can range from a single mismatch to no correct base pairing at all. The free energy difference between the correct and nearly correct base pairs can be as small as 3 kcal/ mol. With complementarity being the only proofreading step, the estimated error frequency would be one wrong amino acid in every 100 amino acids incorporated. However, error frequencies observed in...
11.7K
Quality Assurance
161
Quality assurance is the overarching term used to describe the activities employed to ensure the proper performance of a system. These activities can be classified into three categories: quality control, quality assessment, and internal corrective measures. Typically, these activities work cyclically: quality control is performed before and during the analysis, while quality assessment occurs during and after the investigation. Internal corrective measures are implemented based on the findings...
161
Effects of EDTA on End-Point Detection Methods
310
Different methods, such as visual observance of metal-ion indicators, spectroscopic techniques, and potentiometric methods, can determine the endpoint of an EDTA titration.
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
310
Random and Systematic Errors
11.2K
Scientists always try their best to record measurements with the utmost accuracy and precision. However, sometimes errors do occur. These errors can be random or systematic. Random errors are observed due to the inconsistency or fluctuation in the measurement process, or variations in the quantity itself that is being measured. Such errors fluctuate from being greater than or less than the true value in repeated measurements. Consider a scientist measuring the length of an earthworm using a...
11.2K