使用实验性肺上皮膜和热传感器测量双极烧灼期间的热扩散
Salvatore Siracusano1, Giacomo Marchioro2, Dumitru Scutelnic2
1Department of Life, Health and Environmental Sciences, University of L'Aquila, L'Aquila, Italy.
这项研究调查了在机器人辅助激进前列腺切除术期间,双极电凝的热传播. 调查结果显示,热损伤主要是表面的,为更安全的外科手术技术和设备开发提供了信息.
科学领域:
- 手术技术 手术技术
- 生物医学工程 生物医学工程
- 在瘤学瘤学.
背景情况:
- 节约神经的机器人辅助激进前列腺切除术 (RARP) 通常采用双极电凝.
- 由于神经血管捆的潜在热损伤,双极电凝的使用是有争议的.
研究的目的:
- 在RARP过程中评估组织中的时空空间热分布.
- 为了将热分布与电术诱导的组织损伤相关联.
- 使用富含二氧化碳的环境建模腹腔镜条件.
主要方法:
- 一个密封的玻璃室 (SPC) 模拟了肺上皮膜条件.
- 64个猪肌肉组织 (PMTs) 被用于评估热传播和损伤.
- 一个热摄像头评估了双极性烧伤期间的热分布.
主要成果:
- 30W的双极仪表显示了18mm2的热扩散 (2s) 和28mm2 (4s).
- 在60W时,热扩散为19mm2 (2s) 和21mm2 (4s).
- 组织病理学发现热损伤主要是表面的,而不是深层的.
结论:
- 结果指导了神经节约RARP中准确的双极性体使用.
- 证明了微型热传感器在机器人手术中的可行性.
- 表明开发先进的热内镜器件的潜力.
更多相关视频
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
06:42Plasmonic Photothermal Cancer Therapy: Nanoparticle-embedded Tumor-tissue-mimicking Phantoms for Visualizing Photothermal Temperature Distribution
Published on: May 9, 2025
相关概念视频
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...