新生儿的脑氧和:新生儿和成人NIRS传感器之间的床边比较
Gabriel F T Variane1,2, Alex Dahlen3, Caroline Y Noh4
1Division of Neonatology, Department of Pediatrics, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil. gabriel.variane@pbsf.com.br.
Pediatric research
|June 30, 2023
概括
新生儿传感器显示大脑氧化率 (rScO2) 比NICU婴儿的成年传感器高,但差异不同. 这一发现影响了诊断脑缺氧,并指导了临床实践.
科学领域:
- 新生儿重症监护室新生儿重症监护室
- 医疗器械技术 医疗器械技术
- 监测大脑氧化量的监测
背景情况:
- 新生儿NIRS文献经常使用成人传感器数据来确定目标rScO2范围.
- 新生儿传感器现在是NICU的标准,但与成人传感器相关联的数据有限.
研究的目的:
- 为了比较婴儿中成人和新生儿传感器的同时脑氧化 (rScO2) 测量.
- 确定成人和新生儿NIRS传感器读数之间的相关性和差异.
主要方法:
- 在两个NICU进行前性观察研究 (2019年11月 - 2021年5月).
- 从成人和新生儿传感器收集了6个小时的同时,时间同步的rScO2测量.
- 数据包括各种临床条件下的心率和全身氧和.
主要成果:
- 新生儿传感器在44名婴儿中产生了比成人传感器更高的rScO2读数.
- 在rScO2的差异与绝对值有所不同;在85% (成年人) 的差异约为10%,在55% (成年人) 的读数相似.
- 这种关系被描述为:成年人 = 0.63 × 新生儿 + 18.2.2.
结论:
- 新生儿传感器rScO2通常高于成人传感器rScO2,但差异不是恒定的.
- 假设一个固定的差异可能会导致大脑缺氧的过度诊断.
- 对rScO2的临床解释需要考虑传感器类型和绝对值.
相关概念视频
Special considerations while measuring oxygen saturation
617
Assessing respiratory rate concurrently with pulse measurement is fundamental to patient care, providing valuable insights into the patient's respiratory function. The normal breathing rate for an adult usually falls within a normal range of 12 to 20 breaths per minute. Abnormal respiratory rates can signal underlying health conditions or the need for immediate intervention.
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
617
Pulse Oximetry
364
Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
364