一个具有嵌入式热管理的多通道便携式平台,用于微型介电血液凝固计
IEEE transactions on biomedical circuits and systems
|July 3, 2023
概括
一个带有微流体传感器的新型小型阻抗分析仪 (MIA) 系统准确地测量了血液凝结. 这种便携式平台显示出与已建立的血液静止评估测试在护理地点有很强的相关性.
科学领域:
- 生物医学工程 生物医学工程
- 电气工程 电气工程
- 血液学 血液学 血液学
背景情况:
- 准确的血液静止评估对于患者护理至关重要.
- 目前用于血液凝血测量的方法可能是复杂和耗时的.
- 护理点测试需要便携式和用户友好的设备.
研究的目的:
- 开发和验证一个独立的,多通道的,小型阻抗分析仪 (MIA) 系统用于血液凝血测量.
- 将微流体传感器 (ClotChip) 与MIA系统集成,以进行增强的测量.
- 评估系统的性能与已建立的静血定位测试对比.
主要方法:
- MIA系统使用1MHz的4通道阻抗测量与集成加热.
- 基于树派的计算机处理信号处理和用户界面.
- 该系统使用测试阻抗和体外人体全血样本进行了验证.
主要成果:
- MIA系统与基板分析仪表达了很好的一致性 (rms误差≤0.35%).
- 关键参数T峰和 Δεr,max显示出与ROTEMCT和MCF非常强的正相关性 (分别r=0.98和r=0.92).
- 该系统准确地评估了介电血凝血计参数.
结论:
- 微型阻抗分析仪 (MIA) 系统是血液凝血测量的可行独立平台.
- 该系统显示了在护理点/伤害点 (POC/POI) 进行全面的血液静止评估的巨大潜力.
- 这项技术为快速血液静止评估提供了便携式和准确的解决方案.
相关概念视频
Equipments Used to Measure Body Temperature
1.1K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
1.1K
Capillary Electrophoresis: Instrumentation
268
Capillary electrophoresis instrumentation typically consists of several key components. A high-voltage power supply generates the electric field necessary for the separation by connecting to an anode (the positively charged electrode) and a cathode (the negatively charged electrode) located in buffer reservoirs at each end of the capillary tube. The system includes a sample vial, a fused silica capillary tube coated with polyimide for mechanical strength through which the sample components...
268