代表Elekta MRI-linac财团图像分析工作组提出了改善ADC导出可重复性的建议
Anne L H Bisgaard1, Rick Keesman2, Astrid L H M W van Lier3
1Laboratory of Radiation Physics, Department of Oncology, Odense University Hospital, Kløvervænget 19, 5000 Odense, Denmark; Department of Clinical Research, University of Southern Denmark, J.B. Winsløws Vej 19.3, 5000 Odense Denmark.
概括
看似扩散系数 (ADC) 计算方法显著影响放射治疗反应生物标志物的可重现性. 标准化计算条件对于多中心研究至关重要,以最大限度地减少变化并确保可靠的临床翻译.
科学领域:
- 辐射疗法反应评估评估辐射疗法反应评估
- 医学成像医学成像
- 生物标志物的发展发展.
背景情况:
- 表面扩散系数 (ADC) 是一个有前途的成像生物标志物,用于预测放射治疗反应.
- 对ADC测量的可重现性对于其临床采用至关重要.
- 在ADC划分和计算中的多中心变化可能会阻碍翻译.
研究的目的:
- 评估ADC划分和计算中的多中心变异.
- 确定导致ADC变化的关键因素.
- 在临床研究中提供建议,以尽量减少ADC变异.
主要方法:
- 九个中心分析了各种瘤的相同的MRI数据集.
- 这些中心划分了总瘤体积 (GTV),临床瘤体积 (CTV) 和可活瘤体积 (VTV).
- 使用局部方法和标准化条件 (b值,ROI/基于voxel,平均值/中位数) 计算ADC.
主要成果:
- 计算方法对ADC的可重复性产生了比划界更大的影响.
- 根据ADC的估计,b值集之间的差异为18%,ROI/voxel计算之间的差异为4%.
- 标准化计算条件显著降低了中心间变化 (CVc).
结论:
- 计算方法是多中心研究中ADC变化的重要来源.
- 划分类型 (ROI与voxel) 对可复制性影响较小.
- 建议在多中心研究中对ADC计算协议进行标准化.
相关概念视频
Magnetic Resonance Imaging
5.2K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.2K
NMR Spectrometers: Resolution and Error Correction
729
When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...