交付时间因素对治疗时间和图形疗法计划质量的影响
Takayuki Yagihashi1,2, Tatsuya Inoue3,4, Shintaro Shiba5
1Department of Medical Physics, Shonan Kamakura General Hospital, 1370-1 Okamoto, Kamakura, Kanagawa, 247-8533, Japan.
Scientific reports
|July 27, 2023
概括
在断层疗法规划中的交付时间因子 (DTF) 显著影响治疗效率和计划质量. 优化DTF可以提高计划质量,但持续增加导致更长的治疗时间.
科学领域:
- 医学物理 医学物理
- 辐射瘤学 辐射瘤学
- 放射治疗技术 放射治疗技术
背景情况:
- 雷Station治疗计划系统引入了对断层治疗的交付时间因子 (DTF).
- 了解DTF对治疗效率和计划质量的影响,对于优化放射治疗至关重要.
研究的目的:
- 调查交付时间因子 (DTF) 对断层疗法治疗计划的影响.
- 评估不同的DTF对不同癌症部位的交付效率和计划质量的影响.
主要方法:
- 包括25名头癌,肺癌和前列腺癌患者.
- 螺旋式断层疗法计划被创建并重新计算,DTF范围从1.0到3.0.
- 分析了治疗时间,调制因子,符合性指数,同质性指数和风险器官剂量.
主要成果:
- 治疗时间和调制因子随着DTF增加而单调地增加,每0.1 DTF增量增加约10%
- 计划质量指数 (CI,同质性,OAR剂量) 通常在较高的DTF下得到改善,但肺癌CI除外.
- 计划质量的改善在某些DTF值处停滞不前,而治疗时间继续增加.
结论:
- 交付时间因子 (DTF) 是提高断层疗法计划质量的关键参数.
- 在选择最佳的DTF时,平衡治疗时间和计划质量至关重要.
- DTF优化为图解疗法中改善放射治疗结果提供了一条途径.
更多相关视频
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
15.3K
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
2.8K
相关概念视频
Computed Tomography
4.6K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.6K
Drug Concentration Versus Time Correlation
887
The plasma drug concentration-time curve is a crucial tool in pharmacokinetics, representing the drug's concentration in plasma at different time intervals post-administration. This curve illustrates the drug's journey from absorption into the systemic circulation, distribution to body tissues, and eventual elimination through excretion or biotransformation.
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
887