全参考图像质量评估方法的适用性评估 计算机断层扫描图像的图像
Kohei Ohashi1,2, Yukihiro Nagatani3, Makoto Yoshigoe3
1Division of Health Sciences, Osaka University Graduate School of Medicine, Suita, Japan. kohashi@belle.shiga-med.ac.jp.
Journal of digital imaging
|August 7, 2023
概括
使用全参考IQA (FR-IQA) 方法的客观图像质量评估 (IQA) 适用于计算机断层扫描 (CT) 图像. 视觉信息真实性 (VIF) 与对CT图像质量的人体主观评估的相关性最强.
科学领域:
- 医疗成像医学成像
- 图像质量评估 图像质量评估
- 放射学 放射学是一门学科.
背景情况:
- 准确的图像质量评估 (IQA) 对于医学诊断和治疗至关重要.
- 开发用于自然图像的PSNR和SSIM等全参考IQA (FR-IQA) 方法可能不适合用于计算机断层扫描 (CT) 等医学成像模式.
- 评估现有的FR-IQA方法对CT图像的适用性是必要的,以确保可靠的图像质量评估.
研究的目的:
- 调查客观FR-IQA方法和对CT图像的主观人类评估之间的相关性.
- 确定为自然图像设计的标准FR-IQA方法是否可以可靠地用于CT图像质量评估.
- 与主观评估相比,确定最有效的FR-IQA方法来评估CT图像质量.
主要方法:
- 从六张原始图像中生成了210张扭曲的CT图像,使用噪声和模糊退化.
- 在扭曲的CT图像上评估了九种广泛使用的FR-IQA方法 (PSNR,SSIM,FSIM,IFC,VIF,NQM,VSNR,MSSSIM,IWSSIM).
- 使用双刺激连续质量量表 (DSCQS) 方法对六名观察员进行主观评估,并使用皮尔森线性相关系数 (PLCC),斯皮尔曼等级顺序相关系数 (SROCC) 和根-平均-平方误差 (RMSE) 量化性能.
主要成果:
- 所有九种测试的FR-IQA方法都与主观评估显示出强烈的相关性 (PLCC和SROCC>0.8),表明它们对CT图像的适用性.
- 视觉信息真实性 (VIF) 方法在所有三个指标 (PLCC,SROCC和RMSE) 中表现最好.
- 这些发现表明,VIF是CT图像质量的高度准确的客观测量,与主观的人类评估相美.
结论:
- 标准FR-IQA方法适用于评估CT图像的质量.
- 视觉信息真实性 (VIF) 成为CT图像中最准确的客观IQA指标,与主观的人类感知密切一致.
- 该研究验证了VIF的使用,作为CT成像质量控制和算法开发中的主观评估的可靠替代方案.
相关概念视频
Imaging Studies III: Computed Tomography
28
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
28
Computed Tomography
4.6K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.6K