一种用丝网打印的银电极测定尿样和药品片中的囊的电压计方法
Fang-Ci Kuo1, Chiung-Yao Huang1, Yu-Tung Lin1
1Department of Medical Applied Chemistry, Chung Shan Medical University, Taichung 402, Taiwan.
Journal of pharmaceutical and biomedical analysis
|August 18, 2023
概括
一种新的电压测量方法准确地测量了尿液中的囊水平,有助于囊尿的管理和药品质量控制. 这种快速的技术可在两个样本中提供可靠的囊检测.
科学领域:
- 分析化学 分析化学
- 临床诊断 临床诊断 临床诊断
- 生物医学工程 生物医学工程
背景情况:
- 尿道囊的升高与囊石的形成有关.
- 精确的囊水平的确定对于管理囊性尿症和监测囊药物治疗至关重要.
- 现有的方法可能缺乏及时临床和制药应用所需的速度或效率.
研究的目的:
- 开发一种快速高效的电压计量方法来量化尿囊.
- 为了验证方法的准确性和可靠性,对生物样本和制药产品.
- 为临床诊断囊尿和基于囊的药物的质量控制提供一个有价值的工具.
主要方法:
- 使用商业屏幕打印的银电极建立了电压传感方法.
- 该方法的性能通过校准曲线生成和检测极限确定来评估.
- 对人类尿样和药用囊片进行了测试,以评估康复率.
主要成果:
- 开发的电压计量方法实现了0.65mg/dL的检测极限.
- 对于尿样,观察到89%至109%之间的令人满意的恢复率.
- 对于药物片中的囊,获得了优异的恢复率,从98%到101%不等.
结论:
- 已建立的电压计量方法为确定囊水平提供了快速而准确的手段.
- 该技术适用于囊尿患者的临床诊断和药物配方质量控制.
- 该方法提供了一种有价值的分析工具,用于改善患者管理和确保药物疗效.
相关概念视频
Amperometry: Overview
603
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
603
Potentiometry: Membrane Electrodes
630
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at...
630
Precipitation Titration: Endpoint Detection Methods
1.8K
In argentometric precipitation titrations, endpoints can be detected visually by the Mohr, Volhard, and Fajans methods. In the Mohr method, adding a soluble chromate indicator gives an initial yellow color to the analyte solution. As the titrant is added, the first excess of silver ions forms a red silver chromate precipitate, marking the endpoint. The solution pH should be maintained at about 8 by adding solid CaCO3.
In the Volhard method, a standard excess of AgNO3 is first added to the...
In the Volhard method, a standard excess of AgNO3 is first added to the...
1.8K