openFrame:一个模块化,可持续的开放式显微镜平台,具有单射,双轴光学自动对焦模块,提供高精度和长距离操作
J Lightley1,2, S Kumar1,2, M Q Lim1,3
1Photonics Group, Physics Department, Imperial College London, London, UK.
Journal of microscopy
|August 24, 2023
概括
本研究介绍了openFrame,一个模块化,低成本,开放硬件显微镜平台. 它能够实现可适应,可升级的研究级仪器,并具有用于精确,自动成像的新型自动对焦,包括单分子定位显微镜.
科学领域:
- 显微镜的使用方法
- 光学工程是指光学工程.
- 生物技术是生物技术.
背景情况:
- 传统的显微镜系统可能很昂贵,缺乏模块化.
- 适应显微镜的先进技术,如单分子局部化显微镜 (SMLM) 通常需要专门的,昂贵的设备.
研究的目的:
- 开发一个多功能,低成本,开放硬件显微镜平台 (openFrame).
- 为精确,自动化显微镜设计一个强大的光学自动对焦系统.
- 为了证明平台对SMLM等先进成像技术的实用性.
主要方法:
- 设计了一个模块化,开放的硬件显微镜平台,具有可互换组件.
- 开发了一种新的光学自动对焦,使用直角圆柱形光学来锁定焦点.
- 在openFrame平台上使用多模二极管激光和冷却的CMOS摄像头实现了自动化的SMLM.
主要成果:
- 开放框架平台允许以低成本组装和维护研究级仪器.
- 这种新型自动对焦系统实现了<200nm的精度,失焦率高达±37μm (单拍) 和±68μm (两步).
- 自动化SMLM在经济高效的openFrame系统上成功实现.
结论:
- openFrame提供了一种灵活且负担得起的解决方案,用于构建和升级各种显微镜系统.
- 开发的自动对焦技术显著提高了自动化显微镜的稳定性和准确性.
- 该平台以更容易获取的方式为SMLM等先进研究应用程序提供便利.
相关概念视频
Confocal Fluorescence Microscopy
13.4K
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
13.4K
Imaging Biological Samples with Optical Microscopy
4.8K
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
4.8K
Phase Contrast and Differential Interference Contrast Microscopy
8.2K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
8.2K
Atomic Force Microscopy
3.4K
Atomic force microscopy (AFM) is a type of scanning probe microscopy that can analyze topographic details of various specimens like ceramics, glass, polymers, and biological samples. AFM offers over 1000 times more resolution than the optical imaging system. Images generated from AFM are three-dimensional surface profiles, offering an advantage over the flat, two-dimensional images from other imaging techniques.
The AFM Probe
The probe is regarded as the heart of any AFM setup and comprises the...
The AFM Probe
The probe is regarded as the heart of any AFM setup and comprises the...
3.4K
Super-resolution Fluorescence Microscopy
7.1K
Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
7.1K