通过使用临床上可接受的分析性能规范,验证西门子Atellica®仪器的测量结果
Emmi Rotgers1, Tea Lamberg1, Tero Pihlajamaa1
1Department of Clinical Chemistry, University of Helsinki, and HUS Diagnostic Center, Helsinki and Uusimaa Hospital District, Helsinki, Finland.
Scandinavian journal of clinical and laboratory investigation
|September 6, 2023
概括
临床化学分析仪的性能与临床性能规范进行了验证. 六个测量标准中的三项满足了临床需求,突出了分析变异 (CV) 在设定诊断目标方面的重要性.
科学领域:
- 临床化学和实验室医学
- 分析性能验证 验证 分析性能验证
- 诊断测试评价 诊断测试评价
背景情况:
- 临床化学分析仪的测量需要对临床实用性进行验证.
- 临床上可接受的分析性能规范 (CAAPS) 指导性能标准.
- 个体内生物变异 (CV) 是绩效评估的一个关键组成部分.
研究的目的:
- 为了验证西门子Atellica® Solution化学分析仪的性能.
- 为了将测量的分析变化 (CV) 与六个测量项的建立CAAPS进行比较.
- 根据性能验证,评估分析仪测量的临床适用性.
主要方法:
- 估计分析变化 (CV) 的六个测量:B-HbA1c,U-Alb,P-Na,P-AmylP,P-LDL-C,和P-Crea.
- 使用单个仪器不精确度,并行仪器变化和偏差估计计算的实验CV.
- 对比实验CV与每个测量项的预定义CAAPS.
主要成果:
- 六个测量项中的三项符合各自的CAAPS.
- 报告的特定CVs是:B-HbA1c (1.4%),U-Alb (10.9%),P-Na (1.2%),P-AmylP (8.2%),P-LDL-C (4.9%),P-Crea (4.2%). 据报道,这些CVs中含有B-HbA1c (1.4%),U-Alb (10.9%),P-Na (1.2%),P-AmylP (8.2%),P-LDL-C (4.9%),P-Crea (4.2%). 据报道,这些CVs中含有B-HbA1c (1.4%),U-Alb (10.9%),P-Na (1.2%),P-AmylP (8.2%),P-AmylP (8.2%),P-LDL-C (4.9%),P-Crea (4.2%). 据报道,这些CVs中含有B-HbA1c (1.4%),U-Alb (10.9%),P-Na (1.2%),P-AmylP (8.2%),P-LDL-C (4.9%),P-Crea (4.2%).
- P-Na测量表明需要进行测量改进,特别是在紧急患者样本中.
结论:
- 在确定实验室测试的诊断目标时,必须考虑分析变化 (CV).
- 该研究强调了对临床化学分析仪进行严格性能验证的必要性.
- CAAPS为评估诊断测量的临床适用性提供了关键的基准.
相关概念视频
Instrument Calibration
218
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
218
Data Validation
183
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
Key parameters for method validation include:
183
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
254
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
254
Quality Assurance
155
Quality assurance is the overarching term used to describe the activities employed to ensure the proper performance of a system. These activities can be classified into three categories: quality control, quality assessment, and internal corrective measures. Typically, these activities work cyclically: quality control is performed before and during the analysis, while quality assessment occurs during and after the investigation. Internal corrective measures are implemented based on the findings...
155