基于标准局部一致性干扰测量和基于Scheimpflug的生物仪之间的眼内透镜功率计算的比较:透镜恒定优化的重要性
Sara Sardari1,2, Mehdi Khabazkhoob3, Ebrahim Jafarzadehpur4
1Research and Technology Deputy, Tehran University of Medical Sciences, Tehran, Iran.
Journal of current ophthalmology
|September 8, 2023
概括
IOL Master 500和Pentacam AXL显示了不可互换的眼内透镜 (IOL) 功率计算,尽管平均值差异很小. 广泛的协议限制需要优化Pentacam AXL的镜头常数,以便准确地预测IOL功率.
科学领域:
- 眼科医生 眼科 眼科
- 生物医学光学 生物医学光学
- 光学工程是指光学工程.
背景情况:
- 精确的眼内镜 (IOL) 功率计算对于在白内障手术后实现目标视觉结果至关重要.
- IOL Master 500和Pentacam AXL是用于生物识别和IOL功率计算的现代设备.
- 对比不同生物识别器件和配方的性能对于临床实践至关重要.
研究的目的:
- 使用四种常见公式 (Haigis,Hoffer Q,Holladay 1,SRK/T) 将眼内镜 (IOL) 的功率计算与IOL Master 500和Pentacam AXL进行比较.
- 为了评估两种设备在不同镜头状态之间的一致性.
主要方法:
- 一项涉及1865名60岁以上德黑兰受试者的右眼的横截面研究.
- 使用IOL Master 500和Pentacam AXL进行了生物识别测量和IOL功率计算,使用模式的随机顺序.
- 所有计算都使用了针对Alcon SA60AT镜头的优化ULIB常数和透的目标折射.
主要成果:
- 在所有四种配方中,Pentacam AXL始终计算出比IOL Master 500更高的IOL功率值.
- 平均差异在 -0.33 到 -0.51 D 之间,一致性极限为 95% (例如,Haigis 的 -1.35 到 0.33 D).
- 协议在正常镜头中是最好的,差异与轴长,角质量测量和前腔深度有显著的相关性,角质量测量是最强的预测因素.
结论:
- 虽然IOL Master 500和Pentacam AXL之间的IOL功率的平均差异可能不是临床上显著的,但由于广泛的协议限制,他们的结果是不可互换的.
- 海吉斯配方显示了设备之间特别广泛的协议限制.
- 建议优化Pentacam AXL的镜头常数,以提高IOL Master 500的精度和可互换性.
相关概念视频
Focusing of Light in the Eye
2.9K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
2.9K
Imaging Biological Samples with Optical Microscopy
4.8K
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
4.8K
Phase Contrast and Differential Interference Contrast Microscopy
8.2K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
8.2K