测量集成压力水平的验证耳宽带实耳与合器差异测量
Matthew Urichuk1,2, David Purcell1,2,3, Prudence Allen1,2,3
1Faculty of Health Sciences, School of Communication Sciences and Disorders, Western University, London, Ontario, Canada.
International journal of audiology
|September 18, 2023
概括
使用集成压力水平 (IPL) 测量预测耳模宽带真耳对器差异 (wRECD) 的新方法显示出有希望的结果. 这种技术为传统的探针-麦克风方法提供了可靠的替代品,用于配备助听器.
科学领域:
- 听力学 听力学是指听力学.
- 声学 声学 在声学方面
- 助听器技术 助听器技术
背景情况:
- 精确测量真耳与合器的差异 (wRECD) 对于有效的助听器配件至关重要.
- 对于wRECD,传统的探针-麦克风方法可能是繁的,并且带宽受到限制.
- 集成压力水平 (IPL) 校准传感器为wRECD测量提供了一个潜在的替代方案.
研究的目的:
- 通过使用IPL校准的传感器验证预测耳模 wRECD 的新型测量.
- 评估声学测量管道长度校正对wRECD测量的影响.
- 为了比较建议的预测耳IPL wRECD程序与传统的耳声压水平 (SPL) wRECD测量.
主要方法:
- 22名听力正常的成年人参与了这项研究.
- 单侧耳SPL wRECD和预测耳IPL wRECD用不同的助听器管长度进行测量.
- 声学管长度校正被纳入了IPL wRECD测量程序.
主要成果:
- 探针麦克风和预测的耳IPL wRECD测量在500和2500Hz之间没有发现显著差异.
- 在5000Hz以上,预测的耳IPL wRECD超过了耳SPL wRECDs.
- 与耳型SPL wRECD相比,IPL wRECD测量显示出更高的测试-重新测试可靠性,特别是在更高的频率下.
- 声学管道长度校正有效考虑了耳模声学效应.
结论:
- 基于IPL的wRECD测量是一种可行且有前途的替代探针-麦克风方法.
- 预测的耳形IPL wRECD,测量时没有探针-麦克风放置,显示与SPL wRECDs的良好一致.
- 预计这种方法将扩大wRECD的有效测量带宽,提高助听器的配合精度.
相关概念视频
Measurement of Fluid Pressure
230
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
230
Measurement of Blood Pressure
1.0K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
1.0K
Pressure Gauges
3.5K
Most pressure gauges, like those on scuba tanks, are calibrated to read zero at atmospheric pressure. Readings from such gauges are called the gauge pressure, which is the pressure relative to atmospheric pressure. When the pressure inside the tank exceeds atmospheric pressure, the gauge reports a positive value. Some gauges are designed to measure negative pressure. For example, many physics experiments must take place in a vacuum chamber, a rigid chamber from which some of the air is pumped...
3.5K
Intensity and Pressure of Sound Waves
1.1K
The intensity of sound waves can be related to displacement and pressure amplitudes by using their wave expressions and the definition of intensity. The critical step to achieve this is to write the power delivered by the particles on the wave as the product of force and velocity and simplify the force per unit area as the pressure. The velocity of the medium's particles can be derived from the displacement.
Unlike the time average of a sinusoidal term, which is zero since it is positive...
Unlike the time average of a sinusoidal term, which is zero since it is positive...
1.1K
Special considerations while measuring blood pressure
753
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
753
Assessing Body Temperature - Tympanic membrane
609
Assessing tympanic membrane temperature involves using a tympanic membrane thermometer (TMT). Here is a step-by-step guide:
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
609