开发用于SEM和水溶液样本的直接观测的通用介电常量成像单元
Toshihiko Ogura1, Tomoko Okada1, Michio Hatano2
1Health and Medical Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Central 6, Higashi 1-1-1, Tsukuba, Ibaraki 305-8566, Japan.
概括
研究人员开发了一个新的扫描电子辅助介电显微镜 (SE-ADM) 单元用于标准电子显微镜. 这项创新允许在溶液中直接观察高空间分辨率的标本,克服了以前的限制.
科学领域:
- 材料科学 材料科学 材料科学
- 显微镜的使用方法
- 纳米技术纳米技术
背景情况:
- 电子显微镜需要真空,防止在现场观察溶液中的样品.
- 现有的扫描电子辅助介电显微镜 (SE-ADM) 系统是定制制的,缺乏与扫描电子显微镜 (SEM) 操作系统的集成.
研究的目的:
- 开发一个通用的介电常数成像单元,与商用SEM相兼容.
- 为了使SEM和SE-ADM同时成像以观察溶液中的标本.
主要方法:
- 开发了一个新的SE-ADM单元,可连接到标准的SEM阶段.
- 将SE-ADM介电信号输入集成到SEM的外部终端.
- 在样本持有器中使用了10nm化薄膜来提高分辨率.
主要成果:
- 实现了同时使用 SEM 和 SE-ADM 的成像功能.
- 在观测中显示了4.5nm的空间分辨率.
- 成功观察了培养细胞,PM2.5和溶液中的粘土样本.
结论:
- 开发的通用SE-ADM单元增强了SEM在现场基于解决方案的成像能力.
- 这一进步使得各种标本,包括生物和环境样本,在本地条件下进行详细分析.
相关概念视频
Susceptibility, Permittivity and Dielectric Constant
1.6K
When placed in an external electric field, a dielectric material gets polarized. The charge density in the dielectric material is given by the sum of the bound and free charge densities, while the total charge density can also be written in terms of the total electric field. The bound charge density can be measured in terms of polarization, leading to the relationship between electric displacement and polarization.
1.6K
Dielectric Polarization in a Capacitor
4.7K
The presence of a dielectric medium in a capacitor not only changes the voltage and capacitance but also affects the electric field. In general, dielectrics can be of two types: polar and nonpolar. In a polar dielectric, the positive and negative charges in the molecules are separated by a distance and hence have a permanent dipole moment. In contrast, no such charge separation exists in a nonpolar dielectric, however the nonpolar molecules get polarized in the presence of an external electric...
4.7K
Preparation of Samples for Electron Microscopy
5.5K
To be visualized by an electron microscope, either transmission or scanning, biological samples need to be fixed (stabilized) so the electron beam does not destroy them and dried thoroughly (desiccated/dehydrated) so the vacuum does not affect them. Fixation needs to be done as quickly as possible because the sample properties will start changing as soon as it is removed from its natural environment. For example, in a tissue sample, the oxygen levels begin decreasing, causing an altered...
5.5K