新型基于葡萄糖脱酶的血糖计的系统准确性和干扰评估,用于患者自我测试
Andreas Pfützner1,2,3,4, Daiva Kalasauske1, Mina Hanna1
1Pfützner Science & Health Institute, Mainz, Germany.
Journal of diabetes science and technology
|October 3, 2023
概括
威利恩·纽顿血糖计 (BGM) 符合ISO15197:2015标准,在营销后监测中表现出色. 这种葡萄糖测试系统表现出高精度和最小干扰,确保糖尿病患者可靠的血糖监测.
科学领域:
- 医疗器械 医疗器械
- 临床诊断 临床诊断 临床诊断
- 内分泌学 在内分泌学.
背景情况:
- 欧洲医疗器械法规要求对血糖计 (BGM) 进行营销后监测.
- ISO15197:2015为BGM系统评估提供了性能标准.
- 基于葡萄糖脱酶 (GDH) 的BGM需要仔细评估,因为可能会产生干扰.
研究的目的:
- 根据ISO15197:2015.15的标准来评估Wellion NEWTON BGM的系统性能.
- 在不同受试者群体中评估BGM的准确性和可靠性.
- 为了确定与常见物质的任何潜在干扰.
主要方法:
- 符合ISO15197:2015的系统性能评估.
- 招募了100名受试者 (健康,1型糖尿病,2型糖尿病).
- 通过使用操纵的血液样本对十种选定物质进行了实验室干扰测试.
主要成果:
- 平均绝对相对差异 (MARD) 为4.7%.
- 100%的结果属于共识错误网格区域A (99.7%) 和B (0.3%).
- 仅与已知基于GDH的BGM干扰剂西洛斯 (xylose) 发现干扰.
结论:
- 威利恩·纽顿BGM系统符合ISO15197:2015.15的严格性能要求.
- 在血糖监测方面,BGM表现出高准确度和可靠性.
- 克西洛斯干扰是基于GDH的系统的已知限制,并得到证实.
相关概念视频
Errors occurring during blood pressure monitoring
755
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
755
Amperometry: Overview
600
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...