在床边和专家护士之间使用电子校准器进行计算机化QT和QTc间隔测量的协议
The Journal of cardiovascular nursing
|October 3, 2023
概括
从床边监视器进行的计算机化QT/QTc测量与护士的测量密切一致,提供了一个潜在的更准确和更有效的替代方案. 这一发现表明,计算机化方法可以减少护士的工作量,并提高住院患者心电图监测的准确性.
科学领域:
- 心脏病学 心脏病学
- 临床监测 临床监测
- 电心电图是指心电图.
背景情况:
- 在心电图上延长QT/QTc延长增加了入院患者的Torsade de Pointes风险.
- 手动心电图测量是耗时的,容易出现不准确的情况.
- 自动床边监视器QT/QTc与手动护士测量之间的协议尚未确立.
研究的目的:
- 评估计算机化QT/QTc测量与专家和护士进行的测量之间的一致性.
- 评估任何观察到的QT/QTc测量差异的临床意义.
主要方法:
- 在3个重症监护病房进行的前性观察性研究.
- 使用床边监视器和电子校准器,收集了每位患者最多2次QT/QTc测量.
- 使用布兰德-阿尔特曼测试分析了测量一致性.
主要成果:
- 分析包括34名ICU患者的54次QT/QTc测量.
- 计算机化与专家护士的QT偏差: -11.04±4.45毫秒 (P=.016).
- 计算机化与床边护士的QTc偏差:18.49±7.90毫秒 (P=.022).
结论:
- 计算机和护士执行的QT/QTc测量显示出密切一致.
- 从统计学上显著的差异小于20毫秒,表明没有临床意义.
- 计算机测量可以减少不准确性和护士负担,提高实践遵守.
相关概念视频
Electronic Distance Measuring Instruments
49
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over short...
49
Uncertainty in Measurement: Reading Instruments
38.3K
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
38.3K
Accuracy and Precision
8.9K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. Highly accurate...
8.9K
Distance Corrections
43
To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
43
Detection of Gross Error: The Q Test
6.1K
When one or more data points appear far from the rest of the data, there is a need to determine whether they are outliers and whether they should be eliminated from the data set to ensure an accurate representation of the measured value. In many cases, outliers arise from gross errors (or human errors) and do not accurately reflect the underlying phenomenon. In some cases, however, these apparent outliers reflect true phenomenological differences. In these cases, we can use statistical methods...
6.1K
Measurement of Blood Pressure
1000
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
1000