脉冲波多普勒超声波:采集参数对流量测量的准确性,可变性和影响
Megan K Russ1,2, Nicole M Lafata1,2, Scott H Robertson1,2
1Clinical Imaging Physics Group, Duke University Health System, Durham, North Carolina, USA.
Medical physics
|October 4, 2023
概括
精确的脉冲波多普勒超声波需要持续的质量保证. 这项研究发现,采集参数显著影响流速的准确性,突出了多普勒超声波QA程序中标准化设置的需要.
科学领域:
- 医学成像医学成像
- 超声波技术的超声波技术
- 血管诊断 血管诊断 血管诊断
背景情况:
- 脉冲波多普勒超声波量化血流,对于血管健康评估至关重要.
- 确保多普勒超声波的准确性和一致性对于通过质量保证 (QA) 进行可靠的诊断至关重要.
研究的目的:
- 评估超声波流速测量的准确性,可重现性和扫描仪间的变性.
- 为了确定系统获取参数如何影响使用流体幻影测量流速的准确性.
主要方法:
- 在五个临床超声波系统上使用线性和曲线性传感器与校准的流量幻影进行了测量.
- 分析了峰值和平均流速,通过平均绝对百分比误差来评估扫描器内和扫描器间的变化和准确性.
- 对接收增益,测量深度和光束转向对准确度的影响,与地面真实流速对比进行了检查.
主要成果:
- 对于两种传感器类型的峰值流速测量,观察到具有统计意义的扫描器间变异性,尽管差异很小.
- 平均流速没有显示出统计学上显著的扫描器间变化.
- 接收增益,深度和光束转向影响了线性传感器的准确性,而曲线性传感器的准确性与增益一致,但对深度变化敏感.
结论:
- 标准化的采集和设置参数对于开发可靠的多普勒超声波质量保证程序至关重要.
- 一致的参数选择确保了脉冲波多普勒超声波的可靠和有意义的质量保证.
相关概念视频
Assessing Blood pressure using a doppler ultrasound
1.4K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.4K
Doppler Effect - II
3.4K
The Doppler effect has several practical, real-world applications. For instance, meteorologists use Doppler radars to interpret weather events based on the Doppler effect. Typically, a transmitter emits radio waves at a specific frequency toward the sky from a weather station. The radio waves bounce off the clouds and precipitation and travel back to the weather station. The radio frequency of the waves reflected back to the station appears to decrease if the clouds or precipitation are moving...
3.4K
Equipments Used To Measure Blood Pressure
991
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
991
Doppler Effect - I
3.6K
The Doppler effect and Doppler shift were named after the Austrian physicist and mathematician Christian Johann Doppler in 1842, who conducted experiments with both moving sources and moving observers. Consider an observer standing on a street corner, observing an ambulance with a siren sound passing by at a constant speed. The observer experiences two characteristic changes in the sound of the siren. Initially, the sound increases in loudness as the ambulance approaches and decreases in...
3.6K