相关实验视频
Updated: Jul 14, 2025

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
在前列腺的磁共振成像质量的全球变化
Francesco Giganti1, Alexander Ng1, Aqua Asif1
1From the Departments of Radiology (F.G., L.D., S.P., C.A.), Pathology (A.F.), and Urology (M.E., C.M.M., V.K.), University College London Hospital NHS Foundation Trust, London, United Kingdom; Division of Surgery and Interventional Science, UCL, 43-45 Foley St, 3rd Floor, Charles Bell House, London W1W 7TS, United Kingdom (F.G., A. Ng, A.A., V.W.S.C., A. Nathan, P.K., M.E., C.M.M., C.A., V.K.); Royal Free London, NHS Foundation Trust, London, United Kingdom (A. Ng); British Urology Researchers in Surgical Training (BURST), London, United Kingdom (A. Ng, A.A., V.W.S.C., M.R., A. Nathan, V.K.); UCL Medical School, University College London, London, United Kingdom (M.R.); Department of Urology, Whipps Cross Hospital, Barts Health NHS Trust, London, United Kingdom (P.K.); Centre for Medical Imaging, University College London, London, United Kingdom (S.P.); and National Cancer Imaging Translational Accelerator (NCITA), Division of Medicine, UCL, London, United Kingdom (C.B.G.).
在全球范围内观察到前列腺MRI质量的高变化. 实施前列腺成像质量 (PI-QUAL) 系统和协议修改显著提高了前列腺癌检测MRI扫描质量.
科学领域:
- 放射学 放射学是一门学科.
- 医疗成像医学成像
- 在瘤学瘤学.
背景情况:
- 前列腺MRI质量差异很大,可能会影响前列腺癌检测的准确性.
- 使用MRI±对比度增强 (PRIME) 的前列腺成像试验旨在标准化和改善前列腺MRI质量.
研究的目的:
- 在PRIME试验中使用PI-QUAL系统前性评估前列腺MRI扫描仪质量.
- 为改善MRI协议提供建议,并评估这些变化的影响.
主要方法:
- 来自41个中心的71个扫描仪的355个多参数MRI研究的前性评估 (第一阶段).
- 扫描仪使用PI-QUAL系统被专家泌尿生殖系统放射学家评分.
- 对得分较低的扫描仪提供了反,随后进行了协议修改和重新提交 (第二阶段).
主要成果:
- 在第一阶段,只有32%的扫描仪获得了最佳PI-QUAL评分 (5).
- 动态对比增强 (DCE) 序列显示出与既定标准 (92%) 最显著的偏差.
- 在第二阶段的协议修改后,97%的扫描仪获得了最佳的PI-QUAL分数.
结论:
- 前列腺MRI采集参数存在显著的全球变异性,特别是对于DCE序列.
- PI-QUAL系统和有针对性的协议修改有效地改善前列腺MRI质量.
相关概念视频
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction