在新型连续腹内压力测量系统 (TraumaGuard) 的体外验证
Salar Tayebi1, Robert Wise2,3,4, Ashkan Zarghami1
1Department of Electronics and Informatics, Vrije Universiteit Brussel, 1050 Brussels, Belgium.
Journal of clinical medicine
|October 14, 2023
概括
这项研究验证了TraumaGuard设备的持续腹内压力 (IAP) 监测与黄金标准相比. 创伤护理显示高精度,可以与黄金标准互换使用,即使有空膀.
科学领域:
- 关键护理医学 关键护理医学
- 手术技术 手术技术
- 医疗器械验证 医疗器械验证
背景情况:
- 持续的腹内压力 (CIAP) 监测对于重症患者至关重要,因为腹内高血压的患病率很高.
- 现有的CIAP监控技术具有不同程度的准确性和可靠性.
- 验证像TraumaGuard这样的新设备对于改善患者护理至关重要.
研究的目的:
- 为了在体外环境中验证TraumaGuard CIAP监测设备与黄金标准 (水柱) 相比.
- 为了比较分析不同的CIAP测量技术 (胃内和血管内).
- 为CIAP设备选择提供技术和临床指南.
主要方法:
- 在体外验证五种CIAP设备 (CiMON,Spiegelberg,Serenno,TraumaGuard,Accuryn) 与水柱黄金标准相比.
- 评估温度和膀填充体积对静脉器械的影响.
- 在48小时的连续监测中,对准度,精度,偏差,相关性,一致性和错误网格分析的分析.
主要成果:
- 创伤护理器是唯一一个在空虚的人工膀上运行的设备.
- 静脉内科器件通常显示出更可靠的测量结果;TraumaGuard,Accuryn和Serenno显示出更高的准确性.
- 塞伦诺和TraumaGuard表现出最好的相关性和一致性 (R2=0.98,99.5%一致性),偏差最小,错误治疗风险低.
结论:
- 根据WSACS的指导方针,TraumaGuard设备被验证为可与CIAP测量的黄金标准互换,即使有空膀,也可.
- 静脉内CIAP测量技术似乎比胃内更可靠.
- 建议对动物和人类进行进一步的体内研究,以确认TraumaGuard的临床效用.
相关概念视频
Sites for measuring blood pressure
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
Assessing Blood pressure using a doppler ultrasound
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: