广场多光谱光学显微镜的空间均照明源
İris Çelebi1, Mete Aslan1, M Selim Ünlü1,2,3
1Department of Electrical and Computer Engineering Boston University, Boston, MA, United States of America.
PloS one
|October 18, 2023
概括
我们开发了有效的统一色彩光集成装置 (EUCLID),以改善生物成像照明. 这种具有成本效益的解决方案显著提高了照明均性,将空间偏差降低到大约1%以获得更好的数据提取.
科学领域:
- 光学工程是指光学工程.
- 生物医学成像技术 生物医学成像技术
- 显微镜照明 照明
背景情况:
- 照明均性对于高质量的广场生物成像至关重要.
- 图像系统中的非理想光学元件会导致空间和光谱不均.
- 现有的照明校正方法往往是复杂和昂贵的.
研究的目的:
- 为了引入一个简单,经济高效的照明源,以纠正统一性.
- 为了展示有效的统一色光集成装置 (EUCLID) 设计.
- 评估EUCLID在提高照明均性方面的表现.
主要方法:
- 欧克利德使用分散反射,可调节的空洞腔室来混合光线.
- 该设备修改光源分布,以弥补光学系统的不均性.
- 描述了光合效率和与传统方法相比的一致性.
主要成果:
- 欧克利德 (EUCLID) 显示,照明均度显著改善.
- 在广的视野中实现了大约1%的最大空间偏差.
- 在尼尔森和科勒对齐中有效的性能,用于多光谱成像.
结论:
- 欧克利德为实现照明均性提供了一个简单且具有成本效益的解决方案.
- 该设备提高了广场生物成像中的数据提取质量.
- 提供了一种实用的方法来纠正显微镜系统中的空间不均.
相关概念视频
Imaging Biological Samples with Optical Microscopy
4.8K
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
4.8K
Total Internal Reflection Fluorescence Microscopy
5.8K
Total internal reflection fluorescence microscopy or TIRF is an advanced microscopic technique used to visualize fluorophores in samples close to a solid surface with a higher refractive index, such as a glass coverslip. TIRF only allows fluorophores in proximity to the solid surface to be excited. When light from a medium with a lower refractive index (such as air) hits the glass coverslip at a critical angle, the light undergoes total internal reflection stead of passing through the glass.
5.8K
Confocal Fluorescence Microscopy
13.3K
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
13.3K
Super-resolution Fluorescence Microscopy
7.0K
Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
7.0K
Phase Contrast and Differential Interference Contrast Microscopy
8.1K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
8.1K