在离子流动性光谱-质谱仪中对H2A-H2B体的离子加热的评估
Kazumi Saikusa1,2, Daiki Asakawa3, Sotaro Fuchigami4
1Research Institute for Material and Chemical Measurement, National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan.
Mass spectrometry (Tokyo, Japan)
|October 20, 2023
概括
离子流动性光谱-质谱 (IMS-MS) 的仪器参数影响蛋白质结构. 陷偏差电压,而不是陷碰撞能量,会影响H2A-H2B二度元.
科学领域:
- 分析化学 分析化学
- 生物物理化学 生物物理化学
- 结构生物学 结构生物学
背景情况:
- 离子移动性光谱-质谱 (IMS-MS) 测量离子m/z和碰撞横截面 (CCS).
- 之前的研究表明,H2A-H2B二元体表现出两个CCS形状群体,归因于内在无序的区域.
- 假设这些形状差异与IMS-MS仪器内的离子加热效应有关.
研究的目的:
- 调查IMS-MS中的仪器参数如何影响H2A-H2B二次数的构造群.
- 确定离子加热在产生观察到的结构多样性的特定作用.
- 评估陷偏差电压与陷碰撞能量对离子内部能量和蛋白质构成的影响.
主要方法:
- 通过系统变化的圆电压,陷碰撞能量 (陷CE) 和陷偏差电压分析了H2A-H2B二元体的到达时间分布.
- 使用烯衍生物作为温度敏感探头来量化不同仪器参数所传递的内部能量.
- 仪器参数的相关变化与观察到的形状群体和分析离子的内部能量水平.
主要成果:
- 发现H2A-H2B二元体的两个不同的构造群体主要受到陷偏差电压的影响.
- 对温度敏感的探测器表明,陷CE向离子传递的内部能量比陷偏差电压更多.
- 由陷偏差电压调节的内部能量的微妙变化足以诱导H2A-H2B二度体中的结构多样性.
结论:
- 在IMS-MS中观察到的H2A-H2B二元体的结构多样性对陷偏差电压敏感.
- 虽然陷CE显著增加了离子内部能量,但陷偏差电压对蛋白质构成有更细微的影响.
- 仔细优化陷偏差电压对于准确的生物分子结构分析至关重要,即使其对内部能量的影响似乎很小.
相关概念视频
Chemical Ionization (CI) Mass Spectrometry
769
The molecular ion peak of a molecule in the mass spectrum provides vital information for molecular identification. However, conventional electron impact ionization can lead to the rapid dissociation of some molecular ions before they reach the detector. A milder ionization method is required to increase the lifetime of such ionized analyte molecules. Chemical ionization (CI) is a gas-phase protonation reaction useful for mass-analyzing analyte molecules that are easily protonated to yield the...
769
Electrospray Ionization (ESI) Mass Spectrometry
884
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
884
High-Resolution Mass Spectrometry (HRMS)
1.4K
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
1.4K
Mass Analyzers: Common Types
620
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
620
Mass Spectrometers
5.6K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
5.6K
Mass Spectrometry: Overview
5.3K
Mass spectrometry is an analytical technique used to determine the molecular mass and molecular formula of a compound. The basic principle of mass spectrometry is to generate ions from the analyte molecule and measure these ion abundances against their molecular mass. One common type of ionization, known as electrospray ionization or EI, bombards the analyte molecules in the gas phase with high-energy electron beams. The electron beams displace an electron from the molecule and leave...
5.3K