化学交换和转移 (CEST) 图像的简化评估:局部偏移频率和CEST效应
Daiki Chiba1, Yuki Kanazawa2, Tosiaki Miyati3
1Brain Attack Center Ota Memorial Hospital, 3-6-28 Okinogami, Fukuyama, Hiroshima, 7200825, Japan.
Radiological physics and technology
|October 28, 2023
概括
这项研究引入了一种用于评估化学交换和转移 (CEST) 成像的新幻影. 幻影有效地评估B0场的同质性和CEST效应,这对于优化MRI协议至关重要.
科学领域:
- 磁共振成像技术 磁共振成像技术
- 生物医学工程 生物医学工程
背景情况:
- 化学交换和转移 (CEST) 成像是一种强大的MRI技术.
- 准确评估CEST成像参数,如B0和B1领域的同质性,对于临床应用至关重要.
- 现有的幻影可能无法完全捕捉 in-vivo CEST成像条件的复杂性.
研究的目的:
- 开发和验证一种新的幻影,用于对临床CEST成像设置的全面评估.
- 评估幻影在评估B0和B1场的同质性,CEST对比度和后处理技术中的实用性.
- 通过基于幻影的分析来优化CEST成像协议.
主要方法:
- 构建了一个双切片幻影,其中包括用于局部偏移频率评估的网格部分和用于评估CEST效应的样本部分,其中含有不同度的白蛋白.
- 在3特斯拉MRI扫描仪上进行CEST成像,控制B1幅度 (1.2和1.9μT) 并有/没有B0校正.
- 对获取的图像进行了兴趣区域 (ROI) 分析,比较了不同B1幅度的B0校正和未经校正的CEST图像.
主要成果:
- 在B0校正的Z频谱中,在网格部分,曲线底部向0ppm的转移被证明.
- 在B1 = 1.9μT与1.2μT相比,观察到更广泛的Z频谱曲线形状.
- 对于白蛋白样本,MTRasym值在3.5ppm时,在B1 = 1.9μT与1.2μT时大约是两倍,这表明CEST效应增强.
结论:
- 开发的幻影能够成功地评估和优化B0不均性和CEST效应.
- 幻影作为一个有价值的工具,用于验证CEST成像序列和参数.
- 这种幻影使临床CEST成像协议的改进更加容易,从而提高了诊断准确度.
相关概念视频
Effects of EDTA on End-Point Detection Methods
280
Different methods, such as visual observance of metal-ion indicators, spectroscopic techniques, and potentiometric methods, can determine the endpoint of an EDTA titration.
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
280
NMR Spectroscopy: Chemical Shift Overview
1.5K
The position of the absorption signal of a sample is reported relative to the position of the signal of tetramethylsilane (TMS), which is added as an internal reference while recording spectra. The difference between the absorption frequencies of the sample and TMS (in Hz) is divided by the spectrometer operating frequency (in MHz) to obtain a dimensionless quantity called the chemical shift. It is reported on the δ (delta) scale and expressed in parts per million.
For instance, the proton...
For instance, the proton...
1.5K