在接受重大腹部手术的患者中,Nova StatStrip®葡萄糖计的准确性:一项观察性研究
Keisuke Omiya1,2, Yosuke Nakadate3, Hiroaki Sato4
1Department of Anesthesia, McGill University Health Centre, Royal Victoria Hospital-Glen Site, Montreal, QC, Canada. k.omiya0303@gmail.com.
Canadian journal of anaesthesia = Journal canadien d'anesthesie
|November 3, 2023
概括
诺瓦StatStrip葡萄糖计准确地测量了接受重大腹部手术的患者的血糖. 这种临床检测设备即使在高剂量胰岛素治疗中也符合准确性标准.
科学领域:
- 医疗器械 医疗器械
- 临床化学 临床化学
- 手术瘤学手术瘤学
背景情况:
- 临床检测 (POCT) 设备对于管理重症患者至关重要.
- 诺瓦StatStrip葡萄糖医院仪表系统已被批准用于重症监护中的POCT.
- 在重大腹部手术,特别是肝胆手术过程中,它的准确性仍然没有得到评估.
研究的目的:
- 在接受重大肝胆道手术的患者中评估Nova StatStrip血糖计的准确性.
- 根据已确定的准确性标准评估设备的性能,包括帕克斯错误网格和CLSI指南.
主要方法:
- 一项随机对照试验中的患者对高胰岛素血常糖症 (HNC) 的后期分析.
- 手术前收集了动脉血液样本,并在基线后的1,2,3小时收集了动脉血液样本.
- 使用Nova StatStrip和GEM Premier 5000分析仪测量葡萄糖水平.
主要成果:
- 使用帕克斯错误网格 (ISO15197:2013) 和CLSI POCT12-A3标准来评估准确性.
- 所有135个分析的样本 (包括HNC组的70个) 都在帕克斯错误网格上的可接受区域内.
- 在所有测量时间点上,Nova StatStrip满足了CLSI POCT12-A3准确性标准.
结论:
- 诺瓦StatStrip血糖计在接受重大腹部上部手术的患者中表现出高精度.
- 设备的准确性保持不管在外科手术期间高剂量胰岛素的管理.
相关概念视频
Assessing Blood pressure using a doppler ultrasound
1.4K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.4K
Pre-Procedural Guidelines for Assessing Blood Pressure
573
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
573
Errors occurring during blood pressure monitoring
740
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
740