简单的磁共振温度测量与高率和磁感应度纠正相结合
Sangwoo Kim1, Donghyuk Kim2, Sukhoon Oh3
1Department of Radiological Science, Daewon University College, Jecheon 27135, Republic of Korea.
Bioengineering (Basel, Switzerland)
|November 25, 2023
概括
灵敏度校正的质子共振频率转移 (scPRFS) 提供了准确的MRI温度映射. 这种方法克服了传统的PRFS和光纤传感器的局限性,为非侵入性疗法提供了精确的热监测.
科学领域:
- 磁共振成像 (MRI) 是一种磁共振成像技术.
- 生物医学工程 生物医学工程
- 治疗监测 治疗监测
背景情况:
- 质子共振频率转移 (PRFS) 是一种高分辨率的MRI温度映射技术.
- 用于PRFS验证的光纤热传感器 (FOS) 可以引入由于磁感应变化的温度误差.
- 现有的方法缺乏足够的准确性,无法进行精确的热监测.
研究的目的:
- 开发和验证一个灵敏度校正的PRFS (scPRFS) 方法,用于准确的,高率的温度映射.
- 为了解决传统PRFS和FOS验证在MRI温度计中的局限性.
- 评估scPRFS在非侵入性热疗法中的潜力.
主要方法:
- 实施了基于单个回声的背景删除技术,用于由于磁性敏感性导致的相变校正.
- 开发了scPRFS,用于快速,准确的温度映射.
- 使用幻影和实体实验对FOS进行验证的scPRFS.
主要成果:
- scPRFS表现出与热传感器读数的良好一致.
- 根平均平方误差 (RMSE) 在幻影模型中为0.33-0.36°C,在体外实验中为0.12-0.16°C.
- 实现了高率和温度测量的准确性.
结论:
- scPRFS提供精确的热监测能力.
- 该方法适用于低温和高温的非侵入性疗法.
- scPRFS显示了改善MRI导向热除和监测的巨大潜力.
相关概念视频
NMR Spectrometers: Resolution and Error Correction
700
When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
700
Magnetic Resonance Imaging
5.2K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.2K
Atomic Nuclei: Magnetic Resonance
663
The number of nuclear spins aligned in the lower energy state is slightly greater than those in the higher energy state. In the presence of an external magnetic field, as the spins precess at the Larmor frequency, the excess population results in a net magnetization oriented along the z axis. When a pulse or a short burst of radio waves at the Larmor frequency is applied along the x axis, the coupling of frequencies causes resonance and flips the nuclear spins of the excess population from the...
663