基于袖口的设备校准方法对侵入性和非侵入性大动脉和手臂压力之间的协议的影响
Daniel Bia1, Federico Salazar2, Luis Cinca2
1Departamento de Fisiología, Centro Universitario de Investigación, Innovación y Diagnóstico Arterial (CUiiDARTE), Facultad de Medicina, Universidad de la República, Montevideo, Uruguay.
Clinical physiology and functional imaging
|November 28, 2023
概括
校准方法用于估计大动脉静脉血压 (aoSBP) 影响精度. 不同的方案显示不同的错误和最佳范围,但没有一个方案在评估心脏性质方面明显优于其他方案.
科学领域:
- 心血管生理学心血管生理学
- 生物医学工程 生物医学工程
- 临床测量 临床测量
背景情况:
- 臂袖方法被广泛用于非侵入性大动脉缩血压 (aoSBP) 估计.
- 在准确校准这些手臂测量以反映aoSBP时,仍然存在一些挑战.
研究的目的:
- 评估不同的校准方案如何影响非侵入性aoSBP测量.
- 为了评估校准的非侵入性aoSBP和侵入性测量aoSBP之间的协议.
- 确定校准方法对aoSBP与心脏性质之间的关系的影响.
主要方法:
- 在89名受试者同时进行侵入性和非侵入性血压 (BP) 测量.
- 使用三种校准方案进行非侵入性aoSBP估计:"SD" (腹筋/心筋腕血压),"C" (腹筋腕血压和计算平均血压),以及"Osc" (腹筋腕血压和振荡计衍生平均血压).
- 分析侵入性和非侵入性aoSBP之间的一致性,以及与心声回声学参数的关联.
主要成果:
- 校准方案"C"和"SD"低估了侵入性aoSBP (平均误差为6.9和10.1mmHg),而"Osc"高估了它 (平均误差为-11.4mmHg).
- 在所有方案中,对aoSBP的低估增加了具有更高侵入性的aoSBP水平.
- 每个校准方案都在特定的侵入性aoSBP间隔内显示出最佳的误差范围 ('C': 103-131 mmHg, 'Osc': 159-201 mmHg, 'SD': 101-124 mmHg).
结论:
- 校准方案的选择显著影响非侵入性aoSBP值及其与侵入性测量的一致性.
- 虽然校准方法会影响aoSBP的准确性,但它们对aoSBP与心脏特征之间的关联的影响是相似的.
- 没有单一的校准方法在评估aoSBP与心脏功能之间的关系方面表现出明显的优越性.
相关概念视频
Sites for measruring blood pressure
1.8K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.8K
Assessment of blood pressure in brachial artery(two-step method)
715
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
715
Special considerations while measuring blood pressure
725
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
725
Assessment of blood pressure in brachial artery(one-step method)
599
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
599
Measurement of Blood Pressure
950
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
950
Equipments Used To Measure Blood Pressure
976
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
976