穿戴式耳内脉冲氧计有效地测量了70%至100%之间的氧气和:一项前性协议研究
Catherina Ab Bubb1,2, Michael Weber3, Nadine Kretsch1
1Technical University of Munich, School of Medicine and Health, Department of Anaesthesiology and Intensiv Care Medicine, Munich, Germany.
在健康的成年人中,c-med°α耳内传感器准确地测量氧和度 (SpO2) 和脉率. 这款经过验证的设备符合EN ISO 80601-2-61标准,用于非重症监护病房的监控.
科学领域:
- 医疗器械 医疗器械
- 生理监测 生理监测
- 生物医学工程 生物医学工程
背景情况:
- 在重症监护室以外的术后监测受限于缺乏适当的系统.
- 耳内c-med°α传感器为持续的生理监测提供了一个潜在的解决方案.
研究的目的:
- 为了验证c-med°α耳内传感器测量氧和 (SpO2) 和脉冲率的准确性.
- 根据EN ISO 80601-2-61标准,根据参考脉冲氧计来评估设备的性能.
主要方法:
- 一项前性协议研究,涉及12名健康的成年志愿者 (18-50岁).
- 参与者使用c-med°α传感器和参考指脉冲氧计 (Rad-5) 测量了他们的SpO2和脉冲率.
- 将SpO2水平降低到五个定义的平原 (99%至70%),以评估设备在一系列值的性能.
主要成果:
- c-med°alpha显示了SpO2测量的高精度 (Arms = 1.9%),具有非显著偏差 (-0.1%) 和 -4.0%至3.8%的协议极限.
- 脉冲率测量显示,与参考装置没有显著的偏差 (偏差:0.1分钟-1分钟,一致极限:-3.6到3.7分钟-1分钟,A次数rms:1.8分钟-1分钟).
- 两种设备对SpO2 (0.8%) 的重复系数都相同.
结论:
- 该c-med°α传感器符合EN ISO 80601-2-61标准的要求.
- 该设备足以准确地测量健康成人休息时的SpO2和脉率.
- 这验证了c-med°alpha在非重症监护机构的潜在使用.
更多相关视频
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
相关概念视频
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...