开发和临床验证用于床边血液动力学监测的小型指头探针
Tuukka Panula1, Jukka-Pekka Sirkiä1, Tero Koivisto1
1Department of Computing, University of Turku, Vesilinnantie 5, 20500 Turku, Finland.
iScience
|November 29, 2023
概括
这项研究引入了一种新的手指穿戴设备,用于使用振荡计方法测量无袖式血压 (BP). 该技术提供了方便的门诊监测,还可以评估动脉硬性和检测心房动.
科学领域:
- 生物医学工程 生物医学工程
- 心血管技术的心血管技术
- 医疗器械 医疗器械
背景情况:
- 传统的血压 (BP) 测量依赖于手臂袖子,这可能是门诊监测的繁.
- 需要适合长期使用的方便,非侵入性BP监测技术.
- 脉冲氧计是一种常见的手指穿戴设备,为集成血压测量提供了机会.
研究的目的:
- 开发一个微型,无袖手指探针系统用于血压测量.
- 将血压监测集成到手指穿戴设备中,可能与脉冲氧计相结合.
- 在临床环境中验证开发的设备的准确性和实用性.
主要方法:
- 开发一个微型无袖度指纹探针,使用振荡计方法.
- 使用机动压力机对指尖施加受控的压力,以评估血压.
- 临床验证涉及43名参与者的队列,与手动听觉和自动袖口振荡计进行了比较.
主要成果:
- 该装置显示了精确的静心压和静心压测量,平均误差分别为 (-3.5 ± 8.4) mmHg和 (-4.0 ± 4.4) mmHg.
- 该系统成功评估了18名参与者的动脉硬度.
- 该设备在6名参与者中显示出检测心房动的潜力.
结论:
- 一个小型,方便的手指穿戴设备用于无袖血压测量已经成功开发.
- 该技术使得舒适的门诊监测,使用者最少的不适.
- 该设备对更广泛的心血管健康评估充满希望,包括动脉硬和心房动检测.
相关概念视频
Equipments Used To Measure Blood Pressure
976
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
976
Pre-Procedural Guidelines for Assessing Blood Pressure
569
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
569
Assessing Blood pressure using a doppler ultrasound
1.4K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.4K
Assessment of blood pressure in brachial artery(one-step method)
599
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
599
Sites for measruring blood pressure
1.8K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.8K
Special considerations while measuring blood pressure
725
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
725